Control Solution Test
Chapter 5
Control solution is used to check the performance of the kit.
The meter kit should be check:
• When the meter and strip do not work properly.
• When the test result is unusual or inconsistent.
Control solution range is shown as follow:
Please check your strip vial label for exact range.
G1303190
2018/01
rip BK-U1
7021
5.8-8.6
mg/dL
0.34-0.51
mmol/L
Storage &
Operation
Transportation
4
Perform a Control Test:
Insert an Unused Strip → Make Sure the Code Number is Correct → Shake the
Control Solution Well → Discard First Three Drops → Put One Drop onto a Clean
Surface → Touch the Control Solution with Strip Sample Inlet → Test Result Display
after Count Down → Compare the Result with the Range Listed on Test Strip Vial
Note:
• Do not reuse the test strip.
• Marked open date on new open control solution.
• Do not use the meter if the control test is out of range.
• If the control test keeps result wrongly, please contact an authorized distributor.
• Control solutions are not included. Please contact an authorized distributor for purchasing.
7
Control Range LV1
5.8-8.6
mg/dL
0.34-0.51
mmol/L
EB13120901
EB13120901
2018-01
2018-01