Guidance and Manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or
the user of the device should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)IEC
61000-4-2
Electrical fast
transient/burst IEC
61000-4-4
Surge IEC
61000-4-5
Voltage dips, short
interruptions
variations on power
supply IEC
61000-4-11
IEC 60601 test
Compliance level
level
±6 kV contact
±6 kV contact
±8kV air
±8kV air
±2 kV for power
supply lines ±1
±2 kV for power
kV for
supply lines*1)
input/output lines
±1 kV line to line
±1 kV line to line±2
±2 kV line to
kV line to earth
earth
<5 % UT(>95 %
<5 % UT(>95% dip
dip in UT) for 0.5
in UT) for 0.5 cycle
cycle
40 % UT(60 %
40 % UT(60 % dip
dip in UT) for 5
in UT) for 5 cycles
cycles
70 % UT(30 %
70 % UT(30 % dip
dip in UT) for 25
in UT) for 25
cycles<5 %
cycles <5 %
UT(>95 % dip in
UT(>95% dip in
UT)for 5 sec
UT)for 5 sec
Electromagnetic environment
– guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial and/or
hospital environment. If the user of
this device requires continued
operation power mains
interruptions, it is recommended at
the device be powered from an
uninterruptible supply or battery.
24