FAULT IDENTIFICATION
Switch off the device before any procedure and unplug the power cable from the socket.
PROBLEM
CAUSE
The power cable (A12) has not been
The device does not
inserted properly into the device
work
power socket (A11) or into the mains
socket.
The switch (B1) on the device (B) is set
to "OFF".
The cable connector (B4) has not been
inserted into the 12VDC socket (A5).
The device power cable (B4) has not
been inserted.
The medication or saline solution has
not been inserted into the nebuliser
The nebuliser (B2)
(B2) (bowl).
does not nebulise and/
or dispenses the drug
The nebuliser (B2) has not been prop-
correctly.
erly fitted
The nebuliser nozzle (B2) is clogged
The power cable (A12) has not been insert-
ed properly into the device power socket
(B4) or into the mains socket.
The air tubes (A9) (A10) are not proper-
ly connected to the unit (A) and/or the
PEP connector (C3).
The digital pressure
Air tubes (A9) (A10) are bent and/or
gauge (C) does not
damaged and/or kinked.
light up the LEDs
The ring nut (C1) has not been posi-
and/or does not work
tioned correctly.
properly.
The flow regulator filter (A3) is dirty.
Anti-bacterial filter (A8) clogged or
dirty, bent tube (A9) and/or faulty cir-
cuit.
The electronic circuit does not work
21
SOLUTION
Insert the power cable correctly into the socket
(A11) and into the relative mains socket.
Set the switch (A1) of the device (A) to "ON".
Insert the cable connector (B4) into the socket
(A5).
Insert the device cable (B4).
Pour the correct amount of medication or sa-
line solution into the nebuliser (B2).
Disassemble the nebuliser (B2) and reassemble it
correctly as shown in the diagram on the cover.
Disassemble the nebuliser (B2), remove the
nozzle and perform cleaning operations. Failure
to remove medication deposits from the nebu-
liser affects its efficiency and proper operation.
Strictly comply with the instructions contained
in the cleaning, sanitisation and disinfection
chapter.
Insert the power cable correctly into the socket
(A11) and into the relative mains socket.
Check that the tubes are correctly connected
to the relative sockets of the device (A). See the
connection diagram on the cover.
Unroll the tubes (A9) (A10) and check that they
are not crushed, punctured or bent, if so, re-
place them.
Adjust the PEP ring nut (C1) correctly by match-
ing the holes.
Replace the filter (A3).
Restore the tube (A9), replace the filter (A8)
and/or contact an authorised service centre to
have the electronic circuit replaced.
Contact a FLAEM service centre to replace the
electronic circuit.