5 C O M M I S S I O N I N G A N D O P E R AT I O N

5.9 Heating Up

5.9.1 Display

5
4

5.9.2 Default Settings

The Vaporizer offers three factory preset temperature levels:
Basic temperature: 180 °C (356 °F)
Booster temperature offset: +15 °C (+27 °F)
(relative to the basic temperature) corresponds to 195 °C (383 °F)
Superbooster temperature offset: +15 °C (+27 °F)
(relative to the booster temperature) corresponds to 210 °C (410 °F)
X Basic temperature and booster temperature can be changed individually,
whereas the superbooster temperature is fixed.
X Start inhalation at 180 °C (356 °F).
X If one temperature level is no longer sufficient, change to the next level
as described below.

5.9.3 Switching on

X Press the orange ON/OFF switch for at least half a second.
Ö Vibration and lighting up of the display.
Ö The display indicates the heating process by showing the set and actual
temperatures. This can take up to two minutes.
Ö A double vibration signal indicates that the basic temperature has been
reached.

5.9.4 Base Temperature

X Switch on the Vaporizer.
X Press the plus or minus button.
Ö Permanent change of temperature setting between 40 °C and 210 °C
(104 °F and 410 °F).