Table 4 Guidance and MANUFACTURER's declaration – ELECTROMAGNETIC
IMMUNITY – for ME EQUIPMENT and ME SYSTEMS that are not LIFE-
SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in
IMMUNITY
test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Electrical fast
transient/burst
IEC 61000-4-4
such an environment.
IEC 60601
Compliance
test level
3 Vrms
150 kHz to
3 Vrms
80 MHz
3 V/m
80 MHz to
3 V/m
2.5 GHz
level
environment - guidance
RFvcommunications equipment
should be used no closer to
any part of the device, including
cables, than the recommended
separation distance calculated
from the equation applicable to the
frequency of the transmitter.
Recommended separation
d=1.167√P
d=1.167√P 80MHz to 800MHz
d=2.333√P 800MHz to 2,5GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,
should be less than the compliance
level in each frequency range.
Interference may occur in the
vicinity of equipment
marked with the
following symbol:
29
Electromagnetic
Portable and mobile
distance:
distance in metres (m).
a
b

Publicidad