3. Remove the protective plug from the anti-fouling device cup.
4. Use a toothpick to lift each of the anti-fouling devices out of the holder. If necessary, use needle-
nose pliers to carefully break up the device.
Option
To deploy sensor
To clean or store
sensor
Section 5 General information
This product can expose the user to chemicals with silica, crystalline (airborne particles of respirable size), which
is known to the State of California to cause cancer and birth defects or other reproductive harm. For more
information, go to www.P65Warnings.ca.gov.
Revised editions of this user manual are on the manufacturer's website.
5.1 Service and support
The manufacturer recommends that sensors be sent back to the manufacturer annually to be
cleaned, calibrated, and for standard maintenance.
Refer to the website for FAQs and technical notes, or contact the manufacturer for support at
support@seabird.com. Do the steps below to send a sensor back to the manufacturer.
1. Complete the online Return Merchandise Authorization (RMA) form or contact the manufacturer.
Note: The manufacturer is not responsible for damage to the sensor during return shipment.
2. Remove all batteries from the sensor, if so equipped.
3. Remove all anti-fouling treatments and devices.
Note: The manufacturer will not accept sensors that have been treated with anti-fouling compounds for service
or repair. This includes AF 24173 devices, tri-butyl tin, marine anti-fouling paint, ablative coatings, etc.
4. Use the sensor's original ruggedized shipping case to send the sensor back to the manufacturer.
5. Write the RMA number on the outside of the shipping case and on the packing list.
6. Use 3rd-day air to ship the sensor back to the manufacturer. Do not use ground shipping.
7. The manufacturer will supply all replacement parts and labor and pay to send the sensor back to
the user via 3rd-day air shipping.
5.2 Warranty
Refer to the manufacturer's website for warranty information (seabird.com/warranty).
5.3 AF24173 anti-foulant device
AF24173 Anti-Foulant Devices supplied for user replacement are supplied in polyethylene bags
displaying the following label:
Procedure
Insert new anti-fouling devices into the cup, then install the cap onto the cup. Do not
tighten too tight.
Attach the copper assembly to the sensor again.
Do not insert new anti-fouling devices. Install the protective plug. Make sure to remove
the plug before the next deployment or pressurization of the sensor. Damage to the
conductivity cells can be caused if the plugs are not removed.
W A R N I N G
English 11