Setup and connection
22
7.1 Preparing for commissioning
Before turning on the pump, make sure of the following
points:
Necessary operating requirements
Pump
- Connect all hoses correctly
Pump
- The voltage supply system is consistent with
the details on the type plate of the pump.
- Pump outlet not closed or restricted.
- When operating with gas ballast: When vent-
ing the pump through the air inlet, no explosive
or poisonous mixtures can occur.
Pump
- Data of the voltage supply system are consis-
tent with the details on the type plate of the
power supply.
- Pump outlet not closed or restricted.
Tab.17: Operating requirements for commissioning
Translation of Original Operating and Installation Instruction, KNF
Diaphragm pump N86.18 / N811.18