Adapter-600 V CAT III Power Adapter for use with Models PEL102 and PEL103 ................Cat. #2137.98
The power adapter is used to supply the device when it is used for an extended period. The internal
battery of the PEL remains charged, enabling the device to continue to operate in the event of a power
outage up to 30 minutes.
■ Connect the adapter to the voltage inputs of the PEL.
■ Connect the C7 plug from the adapter to the
■ Connect the voltage inputs to the adapter, following the markings on the adapter. Connect the current
inputs directly to the PEL, per normal operation (see the illustration on page 5).
The L1 and N or L1 and L2 voltage inputs must be connected for the adapter to operate.
The adapter can operate on (50, 60, and 400) Hz networks.
Electrical
Range of use
Max. input voltage
Min. input voltage
Max. output voltage
Frequency
Consumption
Environmental
Operation: (32 to 122) °F (0 to 50) °C and (10 to 85) % RH
Storage: (-4 to 158) °F (-20 to +70) °C and (0 to 95) % RH
Indoor use; Level of pollution: 2; IEC 61010-1 600 V CAT III
Altitude: < 2000 m; Storage: < 10,000 m
Mechanical
Dimensions (L x H x D): (5 x 3.8 x 1.5) in (127 x 97 x 39) mm
Weight: 7 oz (200 g) approximately
Cord: 11 in (30 cm) approximately
Ingress protection:
Drop Test: 1 meter
Electromagnetic compatibility: Emissions and immunity in an industrial setting compliant with
IEC 61326-1
Specifications are subject to change without notice.
This instrument contains no parts that can be replaced by personnel who have not been specially
trained and accredited. Any unauthorized repair or replacement of a part by an equivalent may impair
safety.
Cleaning: Disconnect anything connected to the device. Clean it with a soft cloth moistened with
soapy water. Wipe dry completely.
Repair: For all repairs, please contact the factory or your local distributor.
2
ORDERING INFORMATION
WARNING: Use Only on CAT III Pollution 2 Location.
SPECIFICATIONS
Phase-to-neutral voltage: (100 to 346) V
Phase-to-phase voltage: (100 to 600) V
Permanent: 660 V
85 V
(-15%)
ac
360 V peak
(50, 60, and 400) Hz
100 VA max at (50, 60, and 400) Hz
IP 40 per IEC 60529
IK 04 per IEC 50102
MAINTENANCE
USE
input of the PEL.
; Transient: 690 V
ac
ac
ac
ac
PEL Power Adapter