Cleaning
Clean the cable by wiping it with isopropyl alcohol.
WARNINGS
1. Examine cable before use. Do not use a damaged cable or one with exposed wires.
2. Do not connect the MC10 to any monitor for which the MC10 is not intended for use,
as this could result in electromagnetic interference or susceptibility to or from other
devices within the patient environment.
3. As with all medical equipment, carefully route cables to reduce the possibility of
patient entanglement or strangulation.
CAUTIONS
1. Do not sterilize. This cable is not designed for sterilization, including irradiation, or exposure
to steam or ethylene oxide. Do not immerse the cable in cleaning solutions, disinfecting
solutions, or water. Damage to the cable can result.
2. Electrosurgical interference may compromise performance.
3. When disconnecting the cable from the monitor, always depress the buttons on both sides
of the connector as it is removed from the monitor socket. Failure to do so may result in
damage to the cable connector or to the monitor socket.
4. To avoid damaging the MC10, grasp the connector rather than the cable when discon-
necting it from the sensor or from the monitor.
5. Check cable connections. Failure to properly connect the plugs will result in intermittent
readings or no readings.
Warranty
To obtain information about a warranty for this product, contact Covidien Technical Services or
your local Covidien representative.