Thermo-Anemometer / English
• Large LCD digits at bottom of display for Air Velocity and Air Flow
• Smaller LCD digits at top, right of display for Probe Temperature
•
:Low battery indicator
4.Operation
Connecting the Vane
• The vane plug is inserted in the meter's sensor jack at the top of the meter. The plug and
jack are keyed so that the plug can only fit in the jack one way.
• urn the plug carefully until it lines up with the jack and then firmly push the plug in place.
Do not apply undue force or try to twist the plug side-to-side.
• If the vane is not connected to the meter or if the sensor is defective, the LCD display will
indicate OL in place of a Temperature reading.
Air Velocity Measurements (Single Point)
• Turn on the meter using the ON/OFF
• Press UNITS button to select the desired unit of measure.
NOTE: At power up the meter will display the last unit of
measure previously entered.
• Place the sensor in the air stream. Ensure that the air enters the
vane as indicated by the arrow sticker placed inside the vane.
• View the readings on the LCD Display. The large main LCD
display shows the Air Velocity reading. The upper right LCD sub-
display shows the temperature reading.
Air Flow Measurements (CMM / CFM)
• Turn on the meter using the ON/OFF
• Press the UNITS button to select the desired air flow units:
CMM (cubic meters per minute) or CFM (cubic feet per minute).
NOTE: At power up the meter will display the last unit of
measure previously entered.
2
• To begin entering the area in m
or ft
AREA button until it beeps twice. The leftmost digit of the
bottom display will begin to flash.
• Place the sensor in the air stream. Ensure that the air enters
the vane as indicated by the arrow sticker placed inside the
vane. Refer to the diagram. The large main LCD display shows
the Air Velocity reading.
The upper right LCD sub-display shows the temperature reading.
2
button.
button
2
, press and hold the
The meter has 2 memory locations (1 for CFM and 1 for CMM) that can be used to store
commonly used area sizes that you can recall at anytime.
• Press the AREA button until meter beeps twice. A memory location number will appear in
the top right of the display indicating the memory location.
• Use the UNITS button to change the flashing digit Use the MAX/MIN button to move the
decimal After all of the digits are entered, press and hold the AREA button (until it beeps
twice) to save the area into memory and return to CFM or CMM measuring mode.
Data Hold (Air Velocity/Air Flow)
• While taking measurements, press the HOLD button to freeze the air velocity/air flow
reading.
• The HOLD indicator will appear in the bottom of the LCD display.
• Press HOLD again to return to normal operation
MAX/MIN/AVG Record (Air Velocity/Air Flow)
This allows the user to record and view the highest (MAX), lowest (MIN) and average (AVG)
readings.
• Press the button MAX/MIN button. The MAX indicator and RECORD indicator along with
the Max reading will appear on the LCD display and the meter will begin keeping track of the
MAX, MIN and Average values.
• Press the MAX/MIN button again to view the minimum reading. The MIN indicator along
with the minimum reading will appear on the LCD display.
• Press the MAX/MIN button again to view the average reading. The AVG indicator along
with the average reading will appear on the LCD display.
NOTE: Average recording will stop automatically after 2 hours, and the upper LCD sub-display
will show a OFF (only in the average mode).
• Press the MAX/MIN button again to display current readings. NOTE: the meter will keep
recording MAX/MIN/AVG readings.
• To clear and stop MAX/MIN/AVG recording and return to normal operation, press and hold
the MAX/MIN button until the meter beeps twice.
Thermo-Anemometer / English