PURELAB® CHORUS 1 - ENGLISH
3.3 Replacing the Ultraviolet Lamp LC210
Follow the instructions below.
Note.
The Ultraviolet (UV) lamp should be replaced when:
● UV lamp failure occurs
● UV lamp efficiency decrease affecting water purity.
WARNING
UV-C Radiation is harmful to the eyes and skin. UV Lamp should be operated in the chamber.
It is strongly recommended that during the handling of the UV lamp cut-resistant gloves are worn.
Mercury Hazard, do not break. The lamp contains small amount of mercury.
May cause redness or irritation as a result of contact with skin and / or eyes.
1. Process button off, open right door.
2. Isolate the power supply and turn off the water.
4. Disconnect the lamp connector fitted to
the top of the UV lamp.
CAUTION!
Hold on to the UV lamp carefully to ensure it doesnt break.
Chorus 1 Operator Manual MANU39998 LSC ANR GSC Version 02 3/17 Page 8
PC1LSCXM2 / PC1ANRXM2
3. Unclip UV Lamp LC210 from Chorus.
5. Loosen the 2 sercuring screws at the
top of the UV housing.