3. Specifications
Output signals
Signal type
Current (2-wire)
Voltage (3-wire)
USB
Power supply
The power supply depends on the selected output signal
4 ... 20 mA:
DC 10 ... 30 V
■
DC 0 ... 5 V:
DC 10 ... 30 V
■
DC 0 ... 10 V: DC 14 ... 30 V
■
USB 2.0:
DC 5 V
■
Load in Ω
Current output (2-wire): ≤ (power supply - 10 V) / 0.02 A
■
Voltage output (3-wire): > maximum output signal / 1 mA
■
Ingress protection (per IEC 60529)
Angular connector DIN 175301-803 A: IP 65
■
Circular connector M12 x 1 (4-pin):
■
USB connector type A:
■
Cable outlet:
■
The stated ingress protection only applies when plugged in using mating connectors
that have the appropriate ingress protection.
Permissible temperature ranges
Medium:
0 ... +80 °C
■
Ambient: -20 ... +80 °C
■
Storage: -40 ... +85 °C
■
CE conformity
EMC directive, emission (group 1, class B) and immunity (industrial application)
■
With USB signal output and strong electromagnetic disturbances in the frequency
range 650 ... 750 MHz, the accuracy can be influenced by up to 1 %.
Pressure equipment directive
■
RoHS directive
■
For special model numbers, e.g. HP-20000, please note the specifications stated on the
delivery note.
For further specifications see WIKA data sheet PE 81.53 and the order documentation.
WIKA operating instructions pressure sensor, model HP-2
Signal
4 ... 20 mA
DC 0 ... 5 V
DC 0 ... 10 V
USB 2.0
IP 67
Instrument: IP 67, connector: IP 20
IP 67
EN
9