5.
CLEANING INSTRUCTIONS
Disconnect the Charger from the power source.
Clean the exterior of the charger with a cloth that has been dampened (not dripping) using a
hospital disinfectant. Follow applicable hospital protocols during cleaning and dry
immediately.
6.
DISPOSAL/RECYCLE
In accordance with the European Waste from Electrical and Electronic Equipment (WEEE)
directive, we encourage our customers to recycle this product whenever possible. Disposal
of this unit must be performed in accordance with the applicable local environmental
regulations.
In the US a list of recyclers in your area can be found at: http:/www.eiae.org/.
Please contact our Repair Department to issue a return authorization to return product to
manufacturer at the end of product life.
7.
TROUBLESHOOTING
SYMPTOM
Battery not
charging
Charger LED
indicator not
working
LIT-232
®
CUDA
Surgical
Rev. B
POSSIBLE CAUSES/SOLUTIONS
• Ensure charger power cord is plugged in to the AC power outlet.
• Ensure DC connector is plugged into the back of the charger
• Ensure battery is properly seated in the charger bay.
• If repeated attempts to charge the battery have failed, the battery may be
faulty. Replace the battery.
• The charger may have a faulty power cord and/or power brick. Replace the
power cord and/or power brick.
• Ensure charger power cord is plugged in to the AC power outlet.
• Ensure battery is properly seated in the charger bay.
• The charger may have a faulty power cord and/or power brick. Replace the
power cord and/or power brick.
Page 7 of 48
(English)