7. Interface
Command
:spu <0 ... 9>
:ssl <d | e>
Menu settings
:smm <a | c | m >
:ssw <1 ... 100>
:swm <a | c | m | z l v
| s>
:szc <0.1>
:szi <1 ... 60>
Additional commands
:sce <0.1>
WIKA operating instructions, model CPC2000
Answer function
Pressure unit
0 --> Pa
1 --> hPa
2 --> kPa
3 --> mbar
4 --> bar
5 --> Torr
System language
d = German
e = English
Define the operating mode of the instrument
a --> start automatic mode
c --> start control mode
m --> start Measure mode
Step size
1 ... 100 --> step size from operating the up and down buttons in %
Define the operating mode of the instrument
a --> start automatic mode
c --> start control mode
m --> start Measure mode
(only available in CONTROL and MEASURE modes)
z --> ZERO
(only available in CTRL mode)
l --> LEAK test
l --> back to CTRL mode (like [LEAK] button)
v --> VENT (vent the entire system)
(only available in AUTOMATIC mode)
Zero point adjustment before starting the control mode
0 --> Off
1 --> On
Zero point adjustment interval
1 ... 60 --> time in minutes
Set communication interface echo
0 --> Echo Off
1 --> Echo On
On: Default. Sent command is returned, following the response, the
command will be concluded with "OK".
Off: Only the pure response will be transmitted.
6 --> mmHg
7 --> inHg
8 --> psi
9 --> mmH
O
2
10 --> inH
O
2
EN
33

Publicidad