2. Safety / 3. Specifications

2.6. Special conditions for safe use (X conditions)

Pressure surges must be avoided at all costs. Open the shut-off valves slowly.
Temperature increases due to compression warming absolutely must be taken
into account. In these cases it may be necessary to throttle the rate of change
of pressure or reduce the permissible medium temperature.
The end user must ensure that the instrument is connected to the equipotential
bonding of the end-use application via the process connection.
Avoid handling substances that might react dangerously with the materials of
the instrument.
Avoid handling substances that are liable to spontaneous combustion.

3. Specifications

Pressure limitation
Steady:
Fluctuating:
Overpressure safety
Minimum 1 bar absolute pressure (atmospheric pressure),
in addition 10 x full scale value, max. 25 bar absolute pressure
Temperature effect
When the temperature of the measuring system deviates from the reference
temperature (+20 °C):
max. ±0.8 %/10 K of full scale value
Case ingress protection
IP54 (with liquid filling IP65)
For further specifications see WIKA data sheet PM 05.02 and the order
documentation.
1) For general use, no ATEX requirement
WIKA operating instructions pressure gauge model 5 per ATEX
Full scale value
0.9 x full scale value
(per IEC/EN 60529)
1)