Malfunction
St0p
...
is displayed?
... if, after switching on,
measured results are
transmitted for the first
time and two acoustic
signals are heard?
...if no beeps can be
heard while the wireless
group is being set up,
although the device to be
integrated is switched
on?
... only the "SYS" item is
visible in the rf menu?
... only the "SYS" and
"lrn" items are visible in
the rf menu?
... the "APrt" and "Time"
menu items are not
visible in the rf menu?
Cause/Remedy
While setting up a wireless group:
• The setup of the wireless channels is completed
- Switch on the devices which are to be
integrated in the wireless group (see "Set up
wireless group (Lrn)" on page 68).
• The device was unable to send measured
results to the wireless receiver (seca wireless
printer or PC with seca USB wireless module).
– Check that the device is integrated in the
wireless network.
– Check that the receiver is switched on.
• Nearby RF equipment (e.g. mobile phones) are
interfering with reception.
- Make sure that RF equipment is kept at least
1 metre away from transmitters and receivers
in the seca wireless network.
Note:
If such interference is not eliminated, no
further acoustic warnings are given with
subsequent attempts at transmission.
• The device was not recognised
- Switch devices off and back on again (see
"Set up wireless group (Lrn)" on page 68)
• The wireless module is disabled
- Activate wireless module (see "Activate
wireless module (system)" on page 68) (see
"Activate wireless module (system)" on
page 68)
• The wireless module is activated but no wireless
group set up
- Set up wireless group (see "Set up wireless
group (Lrn)" on page 68).
• No wireless printer logged on in the wireless
group
- Use the "lrn" menu item to log the wireless
printer into the wireless group (see "Set up
wireless group (Lrn)" on page 68).
English