6. Commissioning, operation
Setting
1. Press the MENU button. Confirm the menu item MODE with SELECT and press the
/
button until AUTO appears. Confirm with SELECT.
⇒
For the AUTO mode, the submenu points RANGE, UNIT and STEPS are not
relevant.
2. Go to the SETTINGS menu item with the
SELECT button.
⇒
The submenu item zero point adjustment is not relevant.
3. In the submenu, go to the AUTO mode item with the UP/DOWN buttons and confirm
with SELECT.
RANGE: Pressure range that will be covered
■
UNIT: Pressure unit
■
STEPS: The number of steps which will be taken to increase and decrease the
■
pressure. These increments are calculated by the controller itself.
In the submenu point, CONFIGURATION 1, further settings can be defined for the AUTO
mode.
Setting
CYCLES: Here the number of cycles to be carried out in AUTO mode will be set.
■
T START: Delay time to vent the sensor
■
T STOP: Delay time before reaching the maximum defined pressure
■
T HOLD: Hold time until a new pressure value is reached
■
T PAUSE: Delay time between two cycles
■
AUTO ZERO: Automatic zero point adjustment at the completion of each cycle
■
WIKA operating instructions, model CPC2000
/
buttons and confirm with the
EN
21