PROBLEM
The equipment does
not work
The temperature
inside the equipment
is not low enough (the
compressor operates
continuously)
Water collects at the
bottom part of the
fridge
The equipment makes
too much noise
Troubleshooting
REASON
A gap in the electric system
circuit
The compressor hardly ever
activates
The door doesn't shut tight
or is opened too often
The ambient temperature is
above that which has been
outlined in the product
specification
Air circulation at the back of
the equipment is hampered
The equipment is placed
such that it is subject to
constant sunlight, or another
heat source e.g. Radiator
The contents touch the back
wall of the cabinet
The drain opening is
clogged
The equipment is not level
The equipment is touching
another object
SOLUTION
1. Make sure the plug is properly inserted
into the socket
2. Make sure the socket is not faulty
3. Make sure the power supply cord is not
damaged
1. Check to see if the ambient temperature
is below +10C
1. Rearrange the products so they don't
hamper the door
2. Shorten the time the door is open for
1. Check that the equipment is operating at
an ambient temperature that meets the
product specification
1. Move equipment further away from the
wall
1. Move the equipment to a different place
away from heat sources or direct
sunlight
1. Move the products so they aren't in
contact with the back wall
1. Take the cleaning plug and unclog the
condensate opening
1. Place the equipment on an even surface
or use adjustable feet
1. Reposition the equipment so it is
detached from any other objects