Descargar Imprimir esta página

HEBU medical HBS x-touch 200i Instrucciones Para El Uso página 49

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 92
HBS x-touch 200i
7.11 Operating errors
In the case of an operating error, the following display appears:
The following error messages can occur:
Error - display
Cause of error
Neutral electrode not connected,
NE Error - No 1
not known, or NE is defective
Mono1 - Cut hand/
Terminals: cut foot switch or
FS1 Error
button on handle
Mono1 - Coag
Terminals: coag foot switch or
hand/ FS1 Error
button on handle
If one or more numerical codes are displayed, or if the device is indicating any other recognizable
defects, please refer to service@hebumedical.de
HF current does not flow when the foot switch is actuated. The system only reacts with a
tone signal and the display changes colour.
Please check:
That all terminals have been correctly connected
The settings of the foot switch
Whether the neutral electrode has been recognized
The neutral electrode is not detected or not correctly
Check all terminals. If the neutral electrode is still not recognized, exchange it where appropriate.
It is not possible to switch the machine on or off.
Check whether the power cable has been correctly connected and that the socket is intact. If the
error persists, inform the specialist personnel.
The system works perfectly, but the settings cannot be saved.
Inform the specialist personnel.
All manuals and user guides at all-guides.com
Remedying errors
Check the terminal, touch the NE symbol
on the touchscreen for detection, where
applicable exchange NE and/or
connecting cable.
Check the terminals, check the settings,
exchange accessories if necessary. If
error persists, inform specialist personnel.
EN
49

Publicidad

loading

Este manual también es adecuado para:

Hf 9505