OPEN & READ ALL INSTRUCTIONS
BEFORE OPERATING!
Quick Start Card
Medication Cup Cap
Mouthpiece
Battery Powered Controller
AC/DC Controller
with detachable power cord
with attached power cord
IMPORTANT
ATTENTION
Thank you for purchasing the Aeroneb Go Micropump
Nebulizer from evo Medical Solutions! To optimize the
performance and life expectancy of your Aeroneb Go device,
follow the cleaning instructions carefully as instructed on
page 11 of your Aeroneb Go Instruction Manual.
Retain this card with your device for future reference.
IMPORTANT
CAUTION
As with any mechanical device, this product may become
unusable due to an electrical outage, battery depletion, or
mechanical failure. We recommend that you have spare bat-
teries and have a backup device available to you.
If this product is being used to treat a life threatening
condition, a backup device is necessary.
IMPORTANT CLEANING
INFORMATION FOR ALL MODELS
To optimize the performance and life expectancy of
your Aeroneb Go device, follow the cleaning instructions
carefully as instructed on page 11 of your Aeroneb Go
Instruction Manual.
NOTE: Removal and the improper insertion of the retaining ring
from the underside of the medication cup may cause the unit to leak
or malfunction.
NOTE: Please read all instructions before using. Refer to your
Aeroneb
®
Go Nebulizer Instruction Manual for more complete
cleaning instructions.
To order replacement parts or accessories,visit your local home
medical equipment dealer or pharmacy.
If you have any questions or problems regarding your Aeroneb
nebulizer,contact your local home medical equipment dealer or
pharmacy.
Your Aeroneb
®
Go was tested and inspected at the factory and is a
premium device that has a life expectancy of at least one year (730
treatments) with proper care and cleaning.Lack of proper care and
cleaning,disassembly of the retaining ring or misuse will void the
warranty.
BEFORE
ANY OPERATION
Before you begin any operation, continue to the
Quick Start Instruction Guide read throughly and
follow the operating instructions that apply to
your specific model.
QUICK START INSTRUCTION
TO ASSEMBLE NEBULIZER UNIT
Ventilation Holes
Medication Cup
(with OnQ
®
Aerosol
Generator)
Nebulizer Body
AC/DC Adapter
with attached power cord
TO OPERATE AC/DC POWERED CONTROLLER
TO OPERATE BATTERY POWERED CONTROLLER
NOTE: This model has 2 modes of operation: Battery Powered Controller
only OR AC/DC with Battery Powered Controller.
®
Go
GUIDE BEGINS HERE
STEP 1
Before you begin, run the Medication Cup
under hot water, then tap the cup slightly.
STEP 2
Connect the Medication Cup (containing the
OnQ
®
Aerosol Generator) to the Nebulizer Body.
NOTE: Make sure the Aeroneb Go is unplugged
before you add medication to the Medication
Cup.
STEP 3
Fill the Medication Cup with the prescribed
nebulizer medication. Never pour more than
6 mL into the cup. The continuous line on
the inside surface of the cup indicates the
maximum filling level. (When the medication
cup contains medication, the cup should
remain in an upright position).
CAUTION: To avoid damage to the nebulizer
unit, do not use a syringe with needle.
STEP 4
Securely place Cap on Medication Cup.
STEP 5
Attach Mouthpiece or Mask Connector and
mask to Nebulizer Body. NOTE: Masks are an
optional accessory and may be purchased
evo
through
Medical Solutions or your
Aeroneb
®
Go dealer.
Model #7030
1. Insert the AC Powered Controller cable
connector into the back of the Nebulizer Unit
by sliding the cable straight up the nebulizer
body.
2. Plug the AC Powered Controller into a standard
AC electrical outlet using the appropriate outlet
adapter. The green light on the AC Powered
Controller will light up.
3. Visually verify that aerosol is coming out of the
mouthpiece or mask to indicate that the device
is functioning properly.
4. Place the mouthpiece between your teeth, with
your lips firmly sealed around the mouthpiece.
If you use an aerosol mask, place the mask over
your mouth and nose. Breathe in and out slowly
through your mouth until aerosol formation
stops.Do not cover the ventilation holes during
use. NOTE: Aerosol mask is not provided.
5. Unplug the AC Powered Controller from the
wall to turn it off.Disconnect the cable from the
Nebulizer Unit by sliding the cable straight
down the handset, disassemble Nebulizer Unit
and rinse Nebulizer unit Parts after each use.
Wipe dry the Connector end of the cable after
each use. Clean the nebulizer unit and
associated parts daily. Do NOT remove the
retaining ring under the medication cup.
Model #7090
1.
Turn the Battery Powered Controller upside
down. Slide the battery cover in the direction of
the arrow to remove the lid.Put in three (3) "AA"
disposable alkaline or lithium batteries,
matching them to the plus (+) and minus (-)
markings inside the Battery Powered Controller.
Replace the cover on the Battery Powered
Controller and slide the cover until it locks in
place. You can expect at least three (3) hours of
operation from a new set of disposable
batteries, (up to five (5) hours from high
performance batteries).
2. Insert the nebulizer cable connector into the
back of the Nebulizer Unit. Slide the cable
straight up the nebulizer body.
Continued on next column
Continued from previous column
TO OPERATE BATTERY POWERED CONTROLLER
3. Connect the nebulizer cable to the Aeroneb
Battery Powered Controller.
4. Turn on the unit by pressing the "on/off" button
on the Battery Powered Controller. The green
light on the Battery Powered Controller will
light up. Visually verify that aerosol is coming
out of the mouthpiece or mask to indicate that
the device is functioning properly.
NOTE: (1) If you are using batteries and the green light is flashing, your bat-
teries are low. You may finish your treatment,and then replace the three "AA"
batteries. (2) If the green light only comes on while depressing the button,
immediately replace the three "AA" batteries.
5. Place the mouthpiece between your teeth, with
your lips firmly sealed around the mouth piece.
If you use an aerosol mask, place the mask over
your mouth and nose.Breathe in and out slowly
through your mouth until aerosol formation
stops. Do not cover ventilation holes.
NOTE: Aerosol mask is not provided.
NOTE: (1) The Battery Powered Controller is designed to shut off after 16
minutes. If medication still remains in the cup, press the "on/off" button on
the Battery Powered Controller to continue your treatment. (2) Battery life
is dependent upon not only the number of treatments per day, but also the
length of each treatment. It is recommended that you always have a spare
set of new batteries available to you.
6. Press the "on/off" button on the Battery Powered
Controller again to turn the nebulizer off.
Disconnect the cable from the Nebulizer Unit by
sliding the cable straight down the handset,
disassemble Nebulizer Unit and rinse Nebulizer
Unit Parts after each use. Wipe dry the connector
end of the cable after each use. Clean the
Nebulizer Unit and associated parts daily.
Do NOT remove the retaining ring under the
medication cup.
TO OPERATE BATTERY POWERED CONTROLLER
WITH AN AC/DC ADAPTER
NOTE: This model has 2 modes of operation: Battery Powered Controller
only OR AC/DC with Battery Powered Controller.
1. Plug the AC adapter into a standard AC
electrical outlet, a green light on the adapter
indicates that the AC adapter is turned on.
Plug the AC cable into the Control Module,
indicated
by
Classifications and Markings chart in manual).
Cable
AC Adapter
NOTE: The AC/DC Adapter does not charge batteries.
2. Turn on the unit by pressing the "on/off" button
on the Battery Powered Controller. The green
light on the Battery Powered Controller will light
up. Visually verify that aerosol is coming out of
the mouthpiece or mask to indicate that the
device is functioning properly.
3. Place the mouthpiece between your teeth, with
your lips firmly sealed around the mouthpiece.
If you use an aerosol mask, place the mask over
your mouth and nose. Breathe in and out slowly
through your mouth until aerosol formation
stops. Do not cover the ventilation holes during
use. NOTE: Aerosol mask is not provided.
4. Press the "on/off" button on the Battery
Powered Controller again to turn the nebulizer
off.
5. Disconnect the cable from the Nebulizer Unit
by sliding the cable straight down the handset,
disassemble Nebulizer Unit and rinse Nebulizer
Unit Parts after each use. Wipe dry the
connector end of the cable after each use.
Clean the Nebulizer Unit and associated
parts daily. Do NOT remove the retaining ring
under the medication cup.
Aerogen and Aeroneb are registered
Aerogen (Ireland) Limited,
trademarks of Aerogen, Ireland Ltd.
Galway Business Park,
Dangan,
©2007
evo
Medical Solutions
Galway.
www.aerogen.com
P/N 22262 Rev E
Rx Only
Model #7090
®
Go
Model #7090
the
Input
Symbol
(See

Publicidad

loading

Resumen de contenidos para evo Medical Solutions Aeroneb Go

  • Página 1 Wipe dry the Connector end of the cable after carefully as instructed on page 11 of your Aeroneb Go If you use an aerosol mask, place the mask over each use. Clean the nebulizer unit and your mouth and nose.
  • Página 2 Para optimizar el rendimiento y la duración prevista de la formación de aerosol. No cubra los orificios "on/off" en el controlador a pila. La luz verde en su dispositivo Aeroneb Go, siga cuidadosamente las in- de ventilación durante el uso. NOTA: no se pro controlador encenderá.

Este manual también es adecuado para:

70307090