Descargar Imprimir esta página

Cuda surgical Elite 1000 Manual De Usuario página 10

Ocultar thumbs Ver también para Elite 1000:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 56
Nearby radio or TV units may subject this equipment to radio interference.
To avoid adverse electromagnetic effects, DO NOT operate this equipment near RF
CAUTION
energy equipment.
5.2 CONNECTING THE LIGHT CABLE
Connect the fiberoptic cable to endoscope or other instrument and then plug the cable illuminator
end tip into the appropriate port of the turret on the front panel.
6. OPERATION
6.1 POWERING UP THE LIGHTSOURCE
To operate the illuminator: First attach the illuminator to the instrument or headlamp.
Turn on the rear power switch. Turn on the lamp switch. The fan will start. The indicator on lamp
switch will light.
6.2 LIGHT BRIGHTNESS DISPLAY AND CONTROL
The light brightness adjustment enables the user to obtain a brighter or dimmer illumination of the
object of observation.
Adjust the light intensity by sliding the intensity control knob from left (off) to right (brightest) as
shown by the ramp graphics on the dimmer control.
NOTE
The brightness control operates a progressive mechanical shutter. The lamp is running
at full power if the power switch is on no matter where the brightness control is set.
6.3 LAMP LIFE HOURS DISPLAY
The hour meter shows the total elapsed running time of the unit.
7. CLEANING and DISINFECTION
NOTE
Always disconnect the power cord before cleaning the system and while unit is
drying if wet-wiped.
Disinfection agents should be commercially available cleansers commonly used for disinfection of
electronic equipment cabinets in hospitals, such as ethyl or isopropyl alcohols, disinfecting sprays
containing quaternary ammonium compounds or hydrogen peroxide.
Do not use strongly caustic or acidic cleansers such as "Clorox" hypochlorite bleach, ammonia,
muriatic acid or similar products. Do not use acetone, methyl ethyl ketone or halogenated /
chlorinated hydrocarbon solvents or cleaners containing any of these restricted compounds.
Apply cleaning agents by light spray or dampened towels. Do not pour liquids over the cabinet.
Do not allow frank liquids to enter the cabinet seams or ventilation openings.
®
LIT-217 CUDA
Surgical
Rev. D
Page 10 of 107
(English)

Publicidad

loading