4
Technical data
4.1
Electrical safety
Requirements
4.2
Electrical data
Voltage supply
Current consumption
Power consumption
Overvoltage category
Pollution degree
Protection rating
PoDL class
PoDL classification
Electrical connection
Connection element
M12 plug connector
Version
Protection type
Connecting cable
Version
Bandwidth
Line length
DC loop resistance
Temperature resistance
4.3
Inputs
4.3.1
Measurands
Relative humidity
Measuring range
Accuracy
Typical
Max.
Sampling rate
Temperature
Measuring range
Accuracy
Typical
Max.
Sampling ratio
DIN EN IEC 61010-1:2020
The device must be equipped with an electrical circuit that meets the require-
ments for "Limited-energy circuits".
DC 18 to 30 V, PELV
≤ 100 mA
≤ 1.8 W
II
2
DIN EN 61140, Class III (protective low voltage)
Class 11, IEEE 802.3cg
SCCP or Forced Mode
M12 plug connection
IEC 63171-5
2-pole, shielded
IP67
IEC 61156-13, IEC 61156-14
2-core, copper, shielded
≥ 20 MHz
≤ 500 m (AWG22 at DC 24 V)
< 59 Ω
≥ 80 °C
0 to 95 % RH
±2.0 % RH
±2.5 % RH
1 s
-10 to +60 °C
±0.2 °C
±0.4 °C
1 s
4 Technical data
9