14.4 Error messages on the touch screen
Fault
Error code 1008
Error code 1010
Error code 1022
Error code 1024
Error code 1026
Error code 1100
Error code 1104
Error code -1116
Error code -1117
Error code -1118
Error code -1121
Error code 1153
Error code 1154
2142100008L41 1708V002
OBJ_DOKU-0000002448-001-mehrere.indd 41
Probable cause
Connection interrupted
Temperature of unit too high
Subassembly not initialized
Internal data communication fault
Incorrect acquisition mode
Permitted time for scan process
exceeded
Erasure unit fault
Drive feed blocked
Feed position error
Input unit cover open
Input unit fixing mechanism miss-
ing
Unit fault
Internal data communication fault
Troubleshooting
Solution
Update the firmware
i
Allow the unit to cool down
i
Contact technician
i
Fault in software, update the
i
software if required
Contact technician
i
Switch the unit off and back on
i
again
Update the firmware
i
Darken the room
i
Turn the unit so that no light can
i
fall directly into the insertion slot
Select a different acquisition
i
mode
Inform a service technician
i
Update the firmware
i
Reset the scanning modes
i
to the factory settings via the
unit interface or the Imaging
Software
Contact technician
i
Check the belt drive
i
Check for blockage, remove
i
PSP from unit
Contact technician
i
Replace the erasure unit
i
Remove the blockage
i
Contact technician
i
Contact technician
i
Check the feed (ease of move-
i
ment, drive belts)
Close the cover
i
Deactivate cleaning mode
i
Insert the fixing mechanism
i
Deactivate cleaning mode
i
Switch the unit off and on again
i
Update firmware
i
Switch the unit off and on again
i
Update firmware
i