OPERATING INSTRUCTIONS (Continued):
STEP 3 : Check the oxygen pressure gauge to verify that the oxygen cylinder pressure is within the operating range.
STEP 4: To select the Rest and Active settings on the EVOLUTION MOTION OM-900M conserver (1-6) that are equivalent to
your oxygen prescription, press and hold the REST Selector Button on the Selector Knob until the appropriate LED
display light is illuminated. Next, press and hold the Active Selector Button on the Selector Knob until the appropriate
LED display light is illuminated.
NOTE : To check your current settings, press and release the appropriate Rest or Active Selector Button.
NOTE : The Active setting can only be set equal to or greater than that of the Rest setting. Because the EVOLUTION
MOTION stores the settings from the last time it was used, you only have to complete this step before your first use
or after you have made a change to your normal settings.
STEP 5 :
Ensure cannula is properly attached to the EVOLUTION conserver.
STEP 6 :
Properly position the cannula with the nasal prongs facing upward and insert the prong
into nose. Wrap the cannula tubing over the ears and position in front of body
The conserver will now start to deliver oxygen. The amount of oxygen delivered per pulse is
determined by the setting . A sound may be heard each time the unit delivers a pulse of oxygen.
Adequate saturation will be achieved because of the precise time in the breathing cycle in which the
pulse of oxygen is delivered.
The Active setting should be used during exercise or activity and the Rest setting should be used when inactive or at rest.
The Rest and Active settings should be selected in accordance with a physician's prescription or with a physician's approval.
The motion sensor will select Rest or Active mode in accordance with the below table.
16
TM
TM
(see Fig. F ) .
TM
FIGURE F