Conventions
Warnings and Precautions
Consult users manual for important information.
POSSIBLE EXPLOSION HAZARD. Equipment not suitable for use in the presence of a flammable anes¬
thetic mixture with air or with oxygen or nitrous oxide.
This system should only be used by qualified medical personnel.
WARNING: THERE ISA RISK OF INJURY TO TISSUE FROM EXPOSIJRE TO THE INTENSE ILLUMINA¬
TION. THE RISK OF INJURY IS REDUCED BY USING THE MINIMUM LEVEL OF ILLUMINATION NEC¬
ESSARY, BY MINIMIZING EXPOSURE TIME, AND BY AVOIDING CLOSE STATIONARY VIEWING.
WARNING: The high intensity light at the front of the light source and at the tip of a fiber optic bundle may
give rise to high temperatures. To minimize the risk of injury, avoid contact with tissues.
WARNING: As with any bright light, do not look into the beam. Permanent eye injury may result.
To reduce the chance of direct visual exposure to the light, set the light intensity knob to the minimum
level and plug the fiber optic bundle into the light source before turning the power ON.
When the unit is on, never disconnect the fiber optic bundle from the light source or the accessory from the
end of the fiber bundle. High intensity light from the end of the fiber optic bundle and light source can
ignite flammable materials (i.e. drapes).
WARNING: TOTAL SYSTEM RISK CURRENT - EXCLUDING THE PATIENT APPLIED INSTRUMENT -
SHOULD NOT EXCEED 100 MICROAMPS. AN ISOLATION TRANSFORMER IS REQUIRED IF TOTAL
SYSTEM RISK CURRENT EXCEEDS 100 MICROAMPS WHEN ACCESSORIES ARE INTERCONNECTED.
Grounding reliability is achieved only when connected to hospital-use or hospital-grade receptacles.
Inspect electrical plug and cord routinely. Do not use if damage is discovered.
THE COYER OF THE LIGHT SOURCE SHOULD NEVER BE REMOVED. ELECTRICAL SHOCK HAZARD
DUE TO HIGH INTERNAL VOLTAGE. THERE ARE NO USER SERVICEABLE PARTS INSIDE THIS LIGHT
SOURCE EXCEPT FOR THE LAMP AND FUSE, WHICH ARE ACCESSIBLE THROUGH THE BOTTOM
AND SIDE DOOR PANELS. Refer all service to Welch Allyn.
1-1