NOTE:
• Control solution testing results will be stored into the meter's memory and indicated
with "ctl " icon.
if the user entered the control testing in the mode as described above, the control
results will be marked with the "ctl" and they won't be included in the averages. But, if
the control mode was not entered, then the result will be averaged in with the blood test
results of the user.
• Replace the bottle cap on the control solution bottle immediately after use.
Step 8.
Remove the used test strip from the meter and dispose of it in a biohazard container, and
the meter will turn off automatically after 1.5 minutes.
Perform other levels of control solution testing by repeating the above steps.
IMPORTANT:
• Do not reuse test strips. Test strips are for single use only.
We suggest you should periodically review your technique. You can also compare the
result obtained with your meter to a result obtained using a laboratory method or a well-
maintained and monitored system used by your healthcare provider.