Descargar Imprimir esta página

ACE Instruments DA-7100 Manual De Instrucciones página 10

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 23
Display indications
Display
Meaning
Full battery
Depleted battery: Warning signal when battery is
low. The device is operational, but to ensure correct
results, you should insert new batteries.
BAt
When `BAT´ is displayed, the device is not
operational. Change batteries immediately.
FLo
This is displayed when you have blown wrong or too
weakly into the mouthpiece.
Out
When you have not blown into the mouthpiece within
30 seconds, this is displayed.
Out C
This is displayed when the device was started outside
of the approved operating temperature (0 - 40° C).
Calibration
To maintain the accuracy of the breathalyzer, ACE Instruments recommends a calibration after 500 tests or
approx. every 12 months (depending which incident happens first). For a calibration service you can send
the device to ACE Instruments (see www.alkomat.net). The calibration is not mandatory, but the accuracy of
the breathalyzer will decrease steadily without it.
We recommend sending it as a parcel, so that in cases of loss an inquiry or proof of delivery is possible.
Due to changes in the product, the procedures stated can change. This can take place without prior notice
or adaptation of the operating manual and is no cause for complaint, return or price reduction. The user is
responsible for recognizing discrepancies between the operating manual and the actual handling of this
breathalyzer and must react accordingly.
Battery notices
The device associated with this operating manual contains two AAA Alkaline batteries (1.5 V) in delivery
condition. These batteries are not made for recharging. The use of rechargeable batteries is not possible
with this device and can damage it. If the included batteries are depleted, they must not be thrown into the
household waste. Used batteries may contain harmful substances that can cause environmental or health
damage. Please return all depleted batteries to electrical appliance stores, who collect them for free.
13
Battery exchange
To replace the batteries, push the battery cover on the back of the unit downwards and replace the
Action
discharged batteries with two new ones. Be aware of the polarity when inserting the batteries (+/- pole). Put
the battery compartment cover on the rear panel and press it gently on the device until it clicks. In the EU
None required.
the return of used batteries is mandatory by law. All depleted batteries will be recycled. This way valuable
resources like iron, zinc and nickel can be retrieved and are reusable then. Battery recycling is the easiest
Insert new batteries.
lesson in environmental protection – please support it!
The shown garbage bins mean that batteries must not be thrown into the household waste. The signs
Insert new batteries
below the garbage bins are acronyms for the chemical substances inside the battery:
immediately.
Press the start button
again to repeat the test.
Press the start button
to reactivate the
countdown.
Technical data*
Repeat the test as soon
Sensor
as the correct operating
temperature is reached.
Accuracy
Detection Range
Warm-up Time
Dimension
Weight
Power Supply
Working
Temperature
Storage
Temperature
Calibration
* The technical data are subject to change without prior notice for functional improvements.
Hg: Battery contains Mercury
Cd: Battery contains Cadmium
Pb: Battery contains Lead
Fuel cell sensor
+/- 0.05‰ (per mill) with reference to 1.00‰ (per mill = default setting in the EU)
OR +/- 0.005% BAC with reference to 0.100% BAC (default setting in the USA)
OR +/- 0.025mg/L with reference to 0.50mg/L (default setting in UK)
0.00 up to 5.00‰ (per mill) OR
0.000 up to 0.500% BAC (USA) OR
0.00 up to 2.50mg/L (UK)
Within 20 seconds at 0.50‰ (EU setting), it may vary depending on the value of
per mill which was measured in a test shortly done before
120 x 48 x 20 mm (H x W x D)
93g (batteries included)
2 pieces of 1.5V AAA size Alkaline battery
5°C - 40°C
0°C - 50°C
After 500 tests or approx. every 12 months (depending which incident happens
first). For a calibration service you can send the device to ACE Instruments (see
www.alkomat.net).
14

Publicidad

loading