Warnings
Sievers M500 and M500e
Online TOC Analyzers
Warnings
English
This symbol on the instrument indicates that the
user should refer to the manual for operating
instructions.
The iOS System and vial ports contain sharp
needles designed to pierce the septa of sample
vials. Do not put fingers or inappropriate
materials into the iOS System or vial port.
Water in the iOS system may be hot. Before
inserting a vial into the iOS system after operating
in Online mode, slide the door open and wait 30
seconds to allow sample to completely drain.
Inserting a vial before draining can result in hot
water spray projecting upward out of the iOS
system.
This symbol indicates the protective earth terminal
(ground) for the Analyzer.
This symbol indicates the European Union Directive
2012/19/EU for waste electrical and electronic
equipment (WEEE) requires disposing the Analyzer
separately from standard waste.
To avoid potentially dangerous shock, before opening the
Analyzer, stop any analysis and power off the operating
system. Press the Power Off button in the Analyzer's firmware,
then turn off the power switch and disconnect from the power
supply. Any operation requiring access to the inside of the
Analyzer without turning off power, including installation of
maintenance items, could result in injury.
To protect against accidental exposure to ultra-violet radiation,
do not operate the UV lamp outside of its protective housing.
The UV lamp housing may be hot. Use extreme caution
whenever touching the UV lamp housing.
Risk Group 3 Warning: UV emitted from this product. Avoid eye
and skin exposure to unshielded product.
The glass window in the UV reactor indicates proper lamp
operation and should not be viewed too closely (closer than
6 inches) or for extended periods (greater than 1 minute).
Protective eyewear is recommended.
If this instrument is used in a manner not specified by Veolia,
the protection provided by the instrument may be impaired.
To avoid injury and to ensure accurate analysis results, ensure
that the Analyzer door is closed and latched during normal
operation.
DQS 78000-03 MUL Rev. A
Warnings
This symbol on the instrument indicates that
the product does contain restricted substances
included in China RoHS II. Please refer to the
following table.
Please Note: This product contains a UV Lamp (with Hg) which
meets the exemption defined in Annex III 4(a) of RoHS Directive
2011/65/EC.
To assure safe operation of the instrument, the AC Mains Earth
Ground must be securely connected to the enclosure.
Always stop analysis before turning off the Analyzer.
This is a Safety Class I product. It must be must be attached
to a grounded power source. To avoid damage to the Analyzer
and the chance of injury, make sure the Analyzer is connected
to a power line of the proper voltage.
This is a Class A product. In a domestic environment, this
product may cause electromagnetic interference. Take
adequate measures to correct the interference.
Consult the MSDS that accompanies each standards set to
identify proper disposal techniques.
To avoid potentially dangerous shock, do not touch anything
inside the Analyzer while observing the sample pump during a
visual inspection.
Take extreme care to avoid getting electrical components
wet when performing this procedure.
28
English