Check Content Of Calibration Counter; Technical Data; General Technical Data - Seca 704 Manual De Uso

Ocultar thumbs Ver también para 704:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 138

10.2 Check content of calibration counter

11.TECHNICAL DATA

11.1 General technical data

Dimensions
• Depth
• Width
• Height
Weight
Temperature range
Height of figures
Power supply
These seca scales have been calibrated. Calibration
may only be performed by authorised bodies. To
ensure this, the scales are equipped with a calibration
counter which registers every change in data relevant
to calibration.
If you wish to check that the scales are properly cali-
brated, proceed as follows:
1. Switch the scales off if necessary.
2. Hold down any key and start up the scales.
The current content of the calibration counter will
flash in the display for a few seconds.
3. Compare the output content of the calibration
counter with the figure on the calibration counter
sticker.
The two figures must match for calibration to be valid. If
the sticker and calibration counter do not match, recali-
bration is necessary. Contact your service partner or the
seca customer service department. Should recalibration
be necessary, the seca calibration counter sticker
shown above will be replaced by the adjacent recali-
bration sticker to identify the calibration counter read-
ing. This sticker will be secured with an additional seal
by a person authorised to perform recalibration. The
recalibration sticker is available from the seca customer
service department under the number 14-05-01-886.
Technical data for
seca 704
520 mm
360 mm
930 mm
17 kg
+10° C to +40°C
25 mm
Battery
Power supply unit (optional)
Technical data • 65
English

Publicidad

Tabla de contenido
loading

Tabla de contenido