Fever alarm
The thermometer has a built-in optical indicator to warn you in case of elevated
body temperature. The coloured display shows you the temperature range that is
to be estimated.
Baby temp. mode
Adult temp. mode
< 32,0°C (89,6°F)
< 32,0°C (89,6°F)
32,0°C – 37,2°C
32,0°C – 37,4°C
(89,6°F – 99,0°F)
(89,6°F – 99,3°F)
37,3°C – 37,9°C
37,5°C – 37,9°C
(99,1°F – 100,2°F)
(99,5°F – 100,2°F)
38,0°C – 43,0°C
38,0°C – 43,0°C /
(100,4°F – 109,4°F)
(100,4°F – 109,4°F)
> 43,0°C (109,4°F)
> 43,0°C (109,4°F)
If the device's LED lights up red during temperature measurement, the per-
son's body temperature should be constantly monitored and where there
is uncertainty a doctor's advice should be sought.
Use of the device is in no way a substitute for a doctor's examination. In case of
high fever or unclear measurement results, always seek advice from a doctor.
Measuring object / ambient temperature
If possible, the measurement of liquids should take place from outside the vessel
to avoid any condensation forming on the sensor. The fever alarm is switched off
in this mode. The display is constantly lit green.
The object/ambient temperature mode shows only the exposure temperature
of the surface/room temperature, and does not provide any indication of body
temperature. This measurement mode cannot therefore be used to measure fever.
Set the control to object measurement
rature measurement. Point the thermometer at the desired object and press the
measure button. Read the measured temperature on the display.
Setting functions
Sound mode: select signal tone or mute
1. Make sure that the thermometer is switched off.
2. Press and hold the On/Off button for approx. 12 seconds,
until the display shows „ON" or „OFF".
3. Press the On/Off button again in order to select the mode.
4. Press and hold the On/Off button again for approx. 3 seconds
in order to save your selection.
36+
0 - 36 36+
0 - 36
Display
Measured body
Signal
temperature
tone
Lo
too low
1 x long
green
normal
1 x long
orange
elevated
1 x long
red
high
6 x short
Hi
too high
6 x short
. Follow steps 1 to 3 as for body tempe-
- 18 -
Memory function
1. Make sure that the thermometer is switched off.
2. Press and hold the On/Off button for approx. 5 seconds
36+
3. Press the On/Off button to click through the last 25 stored
4. The thermometer starts with the last measured temperature
5. When the memory is full, the oldest value is automatically
6. The display lights up in the respective display colours of green,
36+
0 - 36 36+
7. To exit from the menu item, let the thermometer rest for
0 - 36
36+
8. To erase the memory, remove the batteries.
9. Press and hold the On/Off button while simultaneously
Change temperature scale
1. Make sure that the thermometer is switched off.
2. Press and hold the On/Off button for approx. 8 seconds
3. Press the On/Off button again in order to choose the units of
4. Press and hold the On/Off button again for approx. 3 seconds
36+
until the display reads „MEM".
values.
at number 1.
deleted and the newest value is saved.
orange and red, as long as it is a body temperature measure-
ment that is being displayed.
approx. 10 seconds until it automatically switches off.
replacing the batteries, until the display shows „- - - -".
until the display shows „C" or „F".
temperature measurement.
in order to save your selection.
- 19 -