• Do not spray detergents-disinfectants directly on the head.
• Do not obstruct the openings next to the fan.
• Do not use the product with accessories that are not specified in this manual.
• Data plate affixed to the rear arm of the device.
• The device is not suitable for installation in explosive atmosphere or rich in
oxygen.
• The articulated arm and head joints enable correct positioning of the light
beam. Do not overload the arms and the joints with shocks on the limit switch.
• The dentist must use disposable safety equipment on lamp handles or guarantee
their sterilisation.
• Unsuitable performance of the device does not compromise patient's safety.
• Divide the materials according to the type (ferrous, rubber, plastic etc.).
Dismantling and disposal of materials should be carried out in accordance with
the standards in force in your country, contacting special waste disposal centres,
if required, that are acknowledged and authorized.
• Do not leave small parts of the equipment unattended or at reach of exposed
persons (children) as they can be hazardous.
• Nominal arm load 5 Kg.
• Do not carry out maintenance interventions or replacements other than those
specified in the manual. Any intervention not specified in this manual may
compromise the safety level provided by the device.
• Carry out exclusively the operations given in this manual; centre for any other
issue, contact the technical support.
• FARO does not allow any change to the product unless specifically authorized
in writing; failure to comply with this rule will nullify the compliance with
safety standards and the warranty.
• The device of the original package can be transported or stored for 15 weeks
if the following conditions are met:
⇒ Room temperature from –20° C to + 70°C
⇒ Relative humidity from 10% to 90%
⇒ Atmospheric pressure from 500 to 1060 mBar
17