Checks And Adjustments; Checking And Adjusting - CST/BERGER 56-DGT2 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 20
HANDLING OF POWER SUPPLY
Removing The Battery Compartment
Push down on the battery compartment latch and pull the top of the battery compartment
away from the instrument. (Fig .21)
Replacing Batteries
1. Push down on the hook to remove the metal battery door. (Fig .22)
2. Install batteries as shown. (Fig .23)
3. Slip the projection shown on the metal battery door into the slot shown and push down on the
door until the hook locks the door in place. (Fig .24)
Mounting Battery Compartment
1. Slip the projection on the bottom of the battery compartment into the slot . (Fig .25)
2. Push the top of the battery compartment toward the housing until the battery compartment
latch clicks into place.

CHECKS AND ADJUSTMENTS

Adjustment Precautions (FIG 26)
When attempting to check and adjust an instrument, corrections must be done in a certain order or
the adjustments will not be correct.
Order of adjustment:
1. Check and adjust the plate vial.
2. Check and adjust the circular vial.
3. Check and adjust the optical plummet.
Always recheck your adjustments. We recommend that any adjustments, other than those shown,
be done by a qualified Instrument repair technician.

Checking And Adjusting

1. The plate vial
A) Place the plate vial parallel with the level screws, A and B. Center the bubble using these two
level screws.
B) Turn the instrument 90° and center the bubble using level screw C. (FIG.27)
14 • DGT2 - DGT10

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

56-dgt10

Tabla de contenido