Ermenrich NG60 Gas Detector
EN
Please carefully read the safety instructions and the user manual before using this product. Keep away from children. Use the device only as
specified in the user manual.
The kit includes: gas detector, USB cable, adapter for charging, aluminum carry case, user manual, and warranty.
Charging the device
This device uses a rechargeable Li-ion battery. Connect the power cable to the device and the DC adapter via a USB plug and connect it
to the AC power supply to charge the device. Fully charge for 6–8 hours. For fast charging, turn the device off. Working battery time is
16–18 hours.
!
Always turn off your device before charging it. Do not charge the device in potentially explosive atmospheres.
Getting started
Press the power button (5) and hold it for 3 seconds to turn the device on/off.
The device will start the self-check: the LED indicators will blink twice, the device will emit a short vibration signal and then will beep
twice. After the self-check, the device will display the preset information for low alarm value, high alarm value, and the detection range
upper limit value for combustable gas, oxygen, hydrogen sulfide and carbon monoxide in sequence.
Settings
Press the power button (5) to enter the menu, to enter a submenu, to confirm the selection, and to save changes. Press the left (4) and
right (6) buttons to scroll through the menu. You can view and set the time and date, and you can view the battery status.
You can set a 4-digit password to prevent the device from accidental operation.
Zero adjustment
You can adjust the zero setting according to the actual detection value.
Enter the zero trim submenu, select the gas type, the value, and unit of measurement. Then, save the adjusted value.
For example, if the displayed value is 3 instead of 0, you can set –0004 and save this value: after that the displayed detection value will be 0.
Alarm records
You can view or clear the previous alarm records in the Alarm Rec (Alarm records) section of the menu.
Alarm records view
To switch between records, press left (4) and right (6) buttons.
Alarm records deletion
To clear the alarm value, press the left button (4) and select OK. To cancel deletion, press the right button (6) and select Quit (Cancel).
Usage
Gas detection mode is set by default. The system returns to the gas detection mode after 5 seconds of idle time. The device will
automatically analyze the gas environment and display the result on the screen.
Normal gas concentration
If the gas concentration is below the low alarm value, the detection result will be displayed on the screen.
Abnormal gas concentration
If the gas concentration is above the alarm value, the LED indicators and the backlight will light on, the device will vibrate and emit
a beeping sound, and the current detection value will be displayed on the screen.
Gas type
Oxygen
Methane
Carbon monoxide
Hydroge sulfide
Combustible gas
Ammonia
Hydrogene
Chlorine gas
Hydrogene chloride
Sulfur dioxide
Nitric oxide
Nitrogene dioxide
Carbon dioxide
Hydrogen phosphide
Hydrogen cyanide
Epoxy acetylene
Detection range
Optional range
0–30% vol.
0–30% vol.
0–4% CH4
0–4% CH4
0–1000ppm
0–2000/5000ppm
0–100ppm
0–50/200/1000ppm
0–100% LEL
0–100% LEL
0–100ppm
0–50/500/1000ppm
0–1000ppm
0–40000ppm
0–20ppm
0–100/150ppm
0–20ppm
0–20/150ppm
0–20ppm
0–50/100ppm
0–250ppm
0–500/1000ppm
0–20ppm
0–50ppm
0–5000ppm
0–5%/10% vol. (infrared)
0–20ppm
0–20ppm
0–20ppm
0–20ppm
0–100ppm
0–100ppm
Low alarm
19.5
1.00
50
10
20
20
200
5
5
5
20
5
1000/0.2
5
10
20
3
High alarm
Resolution
23.5
0.1% vol.
2.50
0.01% CH4
200
1ppm
20
1/0.1ppm
50
1% LEL
50
1/0.1ppm
500
1/0.1ppm
10
1/0.1ppm
10
1/0.1ppm
10
1/0.1ppm
125
1/0.1ppm
10
1/0.1ppm
2000/0.5
1ppm/0.1% vol.
10
1ppm
25
1ppm
50
1ppm