1. Safety Information
Use this instrument only as specified in this manual.
Otherwise, the protection provided by the instrument may be damaged.
Refer to the safety information in Warnings and Cautions.
1.1 Safety
Operate dry-blocks in an ambient temperature between +10 and +30°C (+15 to
25°C for optimum accuracy) and a humidity between 5 to 95%(Non condensing).
The fan runs continuously to moderate the internal unit temperature. Always
ensure that the air vents and fan aperture are clear and have at least 150mm
distance between them and any obstructions. NEVEN cover the unit while in
operation, just operate when the fan stops. If the equipment is used in a manner
not specified by the manufacturer, the protection provided by the equipment may
be damaged.
The calibrator can operate at very high temperature. Precautions MUST be taken
to prevent personal injury or damage to surrounding objests. Probes may be hot
when removed from the units and should be placed on a heat resistant surface.
The unit may remain hot for several minutes after switching off. DO NOT swith
off if temperatures above +100°C. Allow to cool before storage.
Caution must be taken when removing probes and inserts from the units. If the
unit is set at a high temperature the inserts will be very hot and could cause burns
to the hands. Please use the insert removal tool supplied. After removal place hot
probes and inserts only on a suitable heatproof surface.
Only palce temperature probes in dry-block holes. These units are designed to be
operated dry. DO NOT sprinkle any liquid into any of the dry-block holes.
These dry-blocks are designed to be rugged and durable but do contain electronics.
DO NOT operate in dirty, dusty or very damp enviroments or near liquids that could
present a hazard from electric shock.
1.2 Cautions
To avoid possible damage to the instrument, follow these guidelines.
DO NOT plug the unit into 220V if the heater switches and fuse holder read
110V. This action will cause the fuses to blow and may damage the instrument.
DO NOT use fluids to clean the target surface.
DO NOT change the values of the calibration constants from the factory
setting values. The correct setting of these parameters is important to the
safety and proper operation of the calibrato.
Dry-well temperature calibrator / English