Descargar Imprimir esta página

Beha-Amprobe ACD-14-PRO Manual Del Usuario página 10

Pantalla doble 600 a trms multímetro de pinza

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 84
MEASUREMENTS
Warning
W
To avoid electric shock or personal injury:
• When measuring current, center the conductor in the clamp.
• Keep fingers behind tactile barrier.
• Use the proper function and range for measurements.
• Disconnect circuit power and discharge all capacitors before testing
resistance, continuity, capacitance or diodes.
• When using probes, keep fingers behind the finger guards.
• Connecting test leads:
- Connect the common (COM) test lead to the circuit before connecting
the live lead;
- After measurement, remove live lead before removing the common
(COM) test lead from the circuit.
Button
SELECT /
HOLD
/
/
P-RMS
Description
Press SELECT button to select the alternative
measurement function on the rotary switch.
Backlight: Press SELECT button > one second to turn
ON LCD backlight. LCD backlight automatically turns
OFF after approximately 20 minutes.
When LCD backlight is ON, press SELECT button > one
second manually turn OFF the backlight.
Press HOLD to freeze the display reading (
displayed) and press HOLD a second time to release
the reading.
W
Warning
To avoid possible electric shock or personal injury,
when Display HOLD is activated, be aware that the
display will not change when you apply a different
voltage.
Press
/
button to toggle between AC A and
Amp-Tip (precise low current mode). For low current
on small diameter wires (< 10mm), the best accuracy is
specified near the jaw tip area.
Press
/
P-RMS button > one second to enter
P-RMS mode ( P-RMS is displayed) to capture inrush
current (80 ms). Press a second time > one second to
exit.
Note: Auto Power Off is automatically disabled under
P-RMS mode.
6
is

Publicidad

loading

Este manual también es adecuado para:

Acd-14-pro-eur