KEEP THESE INSTRUCTIONS
METER CHECK DISK | Code 1705
Do not attempt to separate the components
of the Meter Check Disk (Code 1705). The
Meter Check Disk consists of a disk with a
permanently attached cover.
Do not fi ll the Meter Check Disk with water.
Water is not used during the meter check
procedure.
CALIBRATION OPERATIONS The Meter Check
Disk (MCD) is used to perform two calibration
operations – Check Calibration and Start
Calibration. The Start Calibration procedure
should only be performed if the meter fails the
Check Calibration procedure.
Check Calibration Meters are calibrated at the
time of manufacture. However, it is possible for
the calibration settings to be lost due to power
anomalies or other circumstances. The Meter
Check Disk is used in the Check Calibration
procedure to determine whether the alignment
of the hub and disk are correct. For some meters,
it also evaluates the brightness of the individual
LEDs.
1. Follow the CLEANING procedure to clean the
light chamber and optic lenses.
light chamber and optic lenses.
2. From the main test page, tap
Settings.
3. Tap Calibration.
4. Tap Check Calibration.
5. Remove the Meter Check Disk (Code 1705)
from the foil pouch. DO NOT remove the black
cover from the disk. Insert the Meter Check
Disk (MCD) into the meter and close the lid.
6. Tap Start.
7. The meter will briefl y spin. When it is complete,
six channel values will be displayed. Compare
the displayed channel values to those printed
on the MCD pouch. If the displayed channel
values are within the ranges provided on
the Meter Check Disk pouch, the meter is
calibrated and performing normally. If the
displayed channel values are not within
the ranges provided on the Meter Check
Disk pouch, perform the Start Calibration
procedure.
Note: Range specifi cations are specifi c to the
disk identifi ed by the serial number on the
pouch. The range specifi cations will vary from
disk to disk. The exact readings from a specifi c
disk may vary from meter to meter.
disk may vary from meter to meter.
8. Tap
to return to the test screen.
9. Remove the MCD from the meter and return to
the foil pouch for storage.
Start Calibration The Start Calibration procedure
calibrates the alignment of the hub and disk
for all meters and sets the brightness of the
individual LEDs for compatible meters. Before
performing this calibration procedure, run
the Check Calibration procedure to determine
whether the meter is calibrated and operating
normally. The Start Calibration procedure should
only be completed if the meter fails the Check
Calibration procedure.
1. Follow the CLEANING procedure to clean the
2. From the main test screen, tap
3. Tap Calibration.
4. Tap Start Calibration.
5. Remove the Meter Check Disk (Code 1705)
6. Tap Start.
7. When the calibration is complete the
8. Tap
Angle Calibration checks the alignment of the
hub and disk. The LED Calibration sets the
brightness of the individual LEDs. The results are
reported as pass or fail. If the measurements
pass, the settings will be saved, and the meter
is calibrated. If the analysis fails, contact Support.
to select
Storage Store the disk in a cool, dry place in the
foil pouch with the desiccant. The disk is sensitive
to moisture. Exposure to excessive heat or
moisture may cause degradation of the disk. The
Meter Check Disk contains delicate materials. Do
not drop, jar, or twist the disk. Do not get the disk
wet. Keep the Meter Check Disk clean and dry.
No intente separar los componentes del
Disco de comprobación del medidor (código
1705). El Disco de comprobación del medidor
se compone de un disco con una cubierta
permanente incorporada.
No llene el Disco de comprobación del medidor
de agua. No use agua durante el proceso de
comprobación del medidor.
OPERACIONES DE CALIBRACIÓN El Disco de
comprobación del medidor (MCD) se utiliza
para realizar dos operaciones de calibración:
Comprobar Calibración e Iniciar Calibración.
El procedimiento Iniciar Calibración solo debe
realizarse si falla el procedimiento Comprobar
Calibración del fotómetro.
Comprobar Calibración Los fotómetros se
calibran en el momento de su fabricación.
Sin embargo, es posible que los ajustes de
calibración se pierdan debido a anomalías
eléctricas u otras circunstancias. El Disco de
comprobación del medidor se utiliza en el
light chamber and optic lenses.
Settings.
from the foil pouch. DO NOT remove the black
cover from the disk. Insert the Meter Check
Disk. Close the lid.
message "Angle Calibration Successful" will
appear. Meters that can also perform an
LED calibration will display "LED Calibration
Successful".
Successful".
to return to the test screen.
¡CONSERVE ESTAS
INSTRUCCIONES!
DISCO DE COMPROBACIÓN DEL
MEDIDOR | Código 1705
to select

Publicidad

loading

Resumen de contenidos para LaMotte 1705

  • Página 1 Check Do not attempt to separate the components Calibration procedure. of the Meter Check Disk (Code 1705). The 1. Follow the CLEANING procedure to clean the Meter Check Disk consists of a disk with a light chamber and optic lenses.
  • Página 2 (code disco a otro. Las lecturas exactas de un mismo 1705). Le disque de vérifi cation de photomètre disco pueden variar de un fotómetro a otro. disco pueden variar de un fotómetro a otro.
  • Página 3 Não tente separar os componentes do Disco de 9. Retirer le MCD du compteur et le remettre Verifi cação do Medidor (Código 1705). O Disco dans la pochette en aluminium pour le stocker. de Verifi cação do Medidor é um disco que leva Démarrer l’étalonnage La procédure de...
  • Página 4 Calibração apenas será concluído se o medidor reprovar o procedimento Verifi car a Calibração. LaMOTTE COMPANY 802 Washington Ave • Chestertown • Maryland • 21620 • USA • lamotte.com Software Support: 800-344-3100 x3 • softwaresupport@lamotte.com ©2022 LaMotte Company | 61705 REV1 | 05.23.2022...