Directrices Y Declaración Del Fabricante; Electromagnetic Immunity - .bock belluno Instrucciones De Uso

Tabla de contenido

Publicidad

7
Directrices y declaración del fabricante
Guidance and manufacturer's declaration
– Electromagnetic emission
The nursing care bed is intended for use in the electromagnetic environment specified below.
The customer or the user of the nursing care bed should assure that it is used in such an environment.
Emission test
Complliance
RF emissions
Group 1
CISPR 11 (partly)
RF emissions
Class B
CISPR 11 (partly)
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
Complies
IEC 61000-3-3

– Electromagnetic immunity

The nursing care bed is intended for use in the electromagnetic environment specified below.
The customer or the user of the nursing care bed should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic discharge
± 6 kV contact
(ESD) IEC 61000-4-2
± 8 kV air
Electrostatic transient/burst
± 2 kV for power supply lines
IEC 61000-4-4
± 1 kV for input/output lines
Surge
± 1 kV differential mode
IEC 61000-4-5
± 2 kV common mode
< 5 % U
(>95 % dip in U
Voltage dips, short inter-
40 % U
ruptions and voltage varia-
(60 % dip in U
tions on power supply input
70 % U
lines
(30 % dip in U
IEC 61000-4-11
< 5 % U
(>95 % dip in U
Power frequency (50/60
Hz) magnetic field
3 A/m
IEC 61000-4-8
Note: U
is the a. c. mains voltage prior to application of the test level.
T
Electromagnetic environment - guidance
The medical used bed uses RF energy only for its internal function. Therefore, its RF emis-
sions are very lowand are not likely to cause any interference in nearby electronic equip-
ment.
The nursing care bed is suitable for use in all establishments other than domestic and those
directly connected to the public-voltage power supply network that supplies buildings used
for domestic purpose.
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
< 5 % UT
T
(>95 % dip in UT ) for 0.5 cy-
) for 0.5 cycle
T
cle
T
40 % UT
) for 5 cycles
T
(60 % dip in UT ) for 5 cycles
T
70 % UT
) for 25 cycles
T
(30 % dip in UT ) for 25 cycles
T
< 5 % UT
) for 5 sec
T
(>95 % dip in UT ) for 5 sec
3 A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic mate-
rial, the relative humidity should be at least
30%.
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Mains power quality should be that of a typi-
cal commercial or hospital environment. If the
user of the nursing care bed requires contin-
ued operation during power mains interrup-
tions, it is recommended that the nursing care
bed be powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
61

Publicidad

Tabla de contenido
loading

Tabla de contenido