SETTING UP YOUR OXYGEN CONCENTRATOR
1 .
Position your unit near an electrical outlet in the room where you spend most of your time .
NOTE– Do not connect to an electrical outlet controlled by a wall switch.
WARNING
Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a person utilizing
oxygen therapy. Keep the oxygen concentrator and cannula at least 2 m (6.5 feet) from hot, sparking objects or naked sources of
flame.
2.
Position your unit on a flat surface at least 6 inches (16 cm) from walls, draperies or any other objects that might prevent the proper
flow of air in and out of your oxygen concentrator. The oxygen concentrator should be located in a well-ventilated area to avoid
pollutants or fumes .
NOTE– To move the unit, firmly grasp the handle located on the top of the unit, rolling and/or lifting the unit over pathway obstacles.
3.
Before operating your unit, always check to be sure the air filter (located on the back of your unit) is clean. Proper cleaning of this filter
is discussed in the Caring For Your Concentrator section on page 7 .
4 .
Attach the appropriate oxygen accessories to the oxygen outlet .
Oxygen Tubing Connection:
a .
Thread the supplied oxygen outlet connector onto the oxygen outlet .
b .
Attach the oxygen tubing directly to the connector (Figure 1) .
Oxygen Tubing Connection With Humidification:
If your physician has prescribed an oxygen humidifier as part of your therapy, follow these steps (If using a prefill, go to step b.):
a.
Fill the humidifier bottle as per manufacturer's instructions.
b.
Thread the wing nut located on the top of the humidifier bottle to the oxygen outlet so that it is suspended (Figure 2). Make sure
it is securely tightened .
c.
Attach the oxygen tubing directly to the humidifier bottle outlet fitting (Figure 3).
NOTE– Your physician has prescribed either a nasal cannula or face mask. In most cases, they are already attached to the oxygen tubing.
If not, follow the manufacturer's instructions for attachment.
5 .
Remove the power cord completely from the line cord strap . Make sure the power switch is in the "Off" position and insert the plug
into the wall outlet . The unit is double insulated to guard against electric shock .
NOTE– (only 120 volt units) The plug on the DeVilbiss oxygen concentrator has one blade wider than the other. To reduce the risk of
electric shock, this plug is intended to fit in a wall outlet only one way. Do not attempt to defeat this safety feature. Your healthcare
provider should verify the compatibility of the oxygen concentrator and all of the parts used to connect to the patient before use.
NOTE– To check your oxygen concentrator and accessories for proper operation; 1. Check the output flow by placing the end of the nasal
cannula under the surface of a half-full cup of water and look for the bubbles. 2. Check the system for leaks by bending the nasal prongs
over and squeeze tight to stop the flow of oxygen. Look at the flow meter to see that the indicator ball on the flow meter drops to zero. If
the indicator ball does not drop to zero, check all connections for possible leaks. Parts to check for leaks are: tubing connections,
humidifier bottle and other accessories like firebreaks. Repeat these steps until the flow meter ball drops to zero. Contact your provider or
service supplier immediately if you encounter any problems.
WARNING
Improper use of the power cord and plugs can cause a burn, fire or other electric shock hazards. Do not use the unit if the power cord is damaged.
OPERATING YOUR DEVILBISS OXYGEN CONCENTRATOR
WARNING
Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a person
utilizing oxygen therapy. Keep the oxygen concentrator and cannula at least 2 m (6.5 feet) from hot, sparking objects or naked
sources of flame.
The DeVilbiss 10-liter oxygen concentrators are equipped with a fire mitigating outlet fitting that prevents propagation of fire into
the unit.
In order to prevent a fire propagating from the patient through the cannula towards the unit, a means of protection should be
located as close to the patient as practicable. Please contact your provider for this means of protection.
1 .
Press the power switch to the "On" position . When the unit is turned "On," all three lights (Service Required, Low Oxygen and Normal
Oxygen) on the front panel will illuminate briefly and an audible signal will briefly alarm confirming that the LEDs and audible signal
are functioning properly . The unit will then operate in "start up" mode with the Low Oxygen light lit until a normal oxygen level is achieved, at which time the Normal
Oxygen light will remain lit . The "start up" may take up to 15 minutes .
NOTE– DeVilbiss recommends for optimal service life that the DeVilbiss Oxygen Concentrator be operated for at least 30 minutes after it is powered on. Shorter periods
of operation, operating in extreme temperature/humidity conditions or in the presence of contaminates, and/or handling and storage conditions outside those specified,
may affect the long term reliable operation of the product.
WARNING
Oxygen makes it easier for a fire to start and spread. Do not leave the nasal cannula or mask on bed coverings or chair cushions if the oxygen concentrator
is turned on but not in use, the oxygen will make the materials flammable. Turn the oxygen concentrator off when not in use to prevent oxygen enrichment.
NOTE– If the audible signal alarms but the unit is not operating, there is no power to the unit. Refer to the Minor Troubleshooting chart on page 8 and contact your
DeVilbiss provider if necessary.
NOTE– If an audible low-frequency vibration sound is detected, the unit is not operating properly. Refer to the Minor Troubleshooting chart on page 8, and contact your
DeVilbiss provider if necessary.
2.
Check the flow meter to make sure that the flow meter ball is centered on the line next to the prescribed number of your flow rate.
EN - 6
1
2
3
NO
SMOKING
A-1025