LERBAKKEN 10, 8410 RØNDE DENMARK.
FM Installation drawing 5335Q502
Model 5331C, 5331D, 5333C and 5333D
Hazardous (Classified) Location
Class I,Division1, Groups, A,B,C,D
Class II Division 1 Groups E,F,G or
Class I, Zone 0, IIC
Ambient temperature limits
T4: -40 to + 85 deg. Celcius
T6: -40 to + 60 deg. Celcius
Terminal 1 , 2
Vmax or Ui: 30 V
Imax or Ii: 120 mA
1
Pmax or Pi: 0.84 W
Ci: 1 nF
Li:10 uH
6
Terminal 3,4,5,6
Only passive, or non-energy
5
storing devices such as RTD's
and Thermocouples may be
connected.
SENSOR
Model 5335C, 5335D
Hazardous (Classified) Location
Class I,Division1, Groups, A,B,C,D
Class II Division 1 Groups E,F,G or
Class I, Zone 0, IIC
Ambient temperature limits
T4: -40 to + 85 deg. Celcius
T6: -40 to + 60 deg. Celcius
Terminal 1 , 2
Vmax or Ui: 30 V
Imax or Ii: 120 mA
1
Pmax or Pi: 0.84 W
Ci: 1 nF
Li:10 uH
6
Terminal 3,4,5,6
Vt or Uo: 9.6 V
5
It or Io: 28 mA
Pt or Po: 67.2 mW
Ca or Co: 3.5 uF
SENSOR
La or Lo: 35 mH
Revision date:
Version Revision
2009-09-22
V1R0
5335QE01
WWW.PRELECTRONICS.COM
Non Hazardous Location
Associated Apparatus
or Barrier
+
with
entity Parameters:
UM < 250V
-
Voc or Uo < Vmax or Ui
Isc or Io < Imax or Ii
Po < Pi
Ca or Co > Ci + Ccable
2
La or Lo > Li + Lcable
3
This device must not be
connected to any associated
4
apparatus which uses or
generates more than 250 VRMS
Non Hazardous Location
Associated Apparatus
or Barrier
+
with
entity Parameters:
UM < 250V
-
Voc or Uo < Vmax or Ui
Isc or Io < Imax or Ii
Po < Pi
Ca or Co > Ci + Ccable
2
La or Lo > Li + Lcable
3
This device must not be
connected to any associated
4
apparatus which uses or
generates more than 250 VRMS
5335V112
LERBAKKEN 10, 8410 RØNDE DENMARK.
The entity concept
The Transmitter must be installed according to National Electrical Code
(ANSI-NFPA 70).
When installed in Class II locations the Transmitter shall be installed in an enclosure with a specified
ingress protection of IP6X according to IEC 60529 and dust-tight conduit seals must be used.
Equipment that is FM-approved for intrinsic safety may be connected to barriers based on the ENTITY
CONCEPT. This concept permits interconnection of approved transmitters, meters and other devices in
combinations which have not been specifically examined by FM, provided that the agency's criteria are
met. The combination is then intrinsically safe, if the entity concept is acceptable to the authority having
jurisdiction over the installation.
The entity concept criteria are as follows:
The intrinsically safe devices, other than barriers, must not be a source of power.
The maximum voltage Ui(V
receive and remain intrinsically safe, must be equal to or greater than the voltage (Uo or V
current (Io or I
The sum of the maximum unprotected capacitance (C
interconnecting wiring must be less than the capacitance (C
barrier.
The sum of the maximum unprotected inductance (L
wiring must be less than the inductance (L
The entity parameters Uo,V
manufacturer.
Page:
Revision date:
1/2
2009-09-22
WWW.PRELECTRONICS.COM
) and current Ii(I
), and maximum power Pi(Pmax), which the device can
MAX
MAX
or I
) and the power Po which can be delivered by the barrier.
SC
t
) for each intrinsically device and the
i
) which can be safely connected to the
a
) for each intrinsically device and the interconnecting
i
) which can be safely connected to the barrier.
a
or V
and Io,I
or I
, and C
and L
for barriers are provided by the barrier
OC
t
SC
t
a
a
Version Revision
V1R0
5335V112
5335QE01
or V
) and
OC
t
Page:
2/2