Electromagnetic Compatibility Information
Table 3
Guidance and declaration of m
Nowadays, m any R F wireless equipments have being used in various healthcare
locations
where medical equipment and/or systems
used in close proximity
equipment
and/or systems' basic safety and essential
affected. Arm-type Fully Automatic Digital Blood Pressure Monitor
tested with the immunity test level in the below table and meet the related
requirements
of IEC 60601-1-2:2014. The
keep a minimum distance between RF wireless communications equipment and
this m edical equipment and/or systems as recommended below.
Test
Band
frequency
(MHz)
(MHz)
385
380-390
450
430-470
710
745
704-787
780
810
800-960
870
930
1720
1845
1700-1990
1970
2400-2570
2450
5240
5500
5100-5800
5785
anufacturer-electromagnetic immunity
to medical equipment
Service
Modulation
Pulse
TETRA 4 00
modulation
1 8Hz
FM
GMRS 460
± ٥ kHz deviation
FRS 460
1 kHz sine
LTE
Pulse
Band
modulation
13, 17
217Hz
GSM 8 00/900,
Pulse
TETRA 800,
iDEN 8 20,
modulation
CDMA 850,
18Hz
LTE B and 5
GSM 1800;
CDMA 1 900;
Pulse
GSM 1900;
modulation
DECT;
217Hz
LTE Band 1, 3,
4, 25; UMTS
Bluetooth,WLAN,
Pulse m odulation
802.11 b/g/n,RFID
217Hz
2450,LTE Band 7
Pulse
WLAN
modulation
802.11
217Hz
a/n
are used. When
and/or systems,
performance
customer and/or user should
Maximum
Distance
power
(m)
(W)
0.3
1.8
2
0.3
0.2
0.3
2
0.3
0.3
2
0.3
2
0.3
0.2
164
they are
the medical
may be
has been
help
Immunity
test level
(V/m)
27
28
9
28
28
28
9