Chapter 3 – Working with the titrator TitroLine
6000/7000
In the case of an automatic acceptance of the balance data, the weighed-in quantities will be read in from a
memory. If the memory does not contain any balance data, a message will appear to indicate that no balance
data are present:
Fig. 32
Pressing the Print key will transfer the balance data, too. Titration will then begin directly after the transfer of the
balance data without any further confirmation being necessary. The display will show the measured value (pH,
mV or µA) and the current consumption. The measured value is displayed in a slightly larger font. The top of the
display will show the "Titration is running" status indication and the method being used, i.e. "pH titration":
Fig. 33
Pressing the <Mode>/<F6> will cause the titration curve to be displayed (Fig. 34).
Fig. 34