Before Use; Electrical Installation - Bibby Scientific Stuart SB300 Instrucciones De Funcionamiento

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

English

Before Use

Thank you for purchasing this Bibby
Scientific product. To get the best
performance from the equipment, and for
your own safety, please read these
instructions carefully before use.
If the equipment is not used in the manner
described in this manual the protection
provided may be impaired.
This equipment is designed to operate under the
following conditions:-
❖ For indoor use only
❖ Use in a well ventilated area
❖ Ambient temperature range +5°C to +40°C
❖ Altitude to 2000m
❖ Relative humidity not exceeding 80%
❖ Mains supply fluctuation not exceeding 10%
❖ Overvoltage category II IEC60364-4-443
❖ Pollution degree 2 IEC664
❖ Use with a minimum distance all around of
200mm from walls or other items

Electrical Installation

Before connection please read and
understand this instruction manual and
ensure that the line supply corresponds to
that shown on the rating plate. All models
require a supply rated at: 230V, 50Hz, ~,
single phase.
Power consumption is:
Model
Power
SB300
600W
SB301
50W
SB302
650W
SB500
1500W
SD300
600W
SD500
1500W
THIS INSTRUMENT MUST BE EARTHED
All models are supplied with two mains leads
fitted with IEC plugs for connection to the
instrument. One has a U.K. 3 pin plug and the
other has a 2 pin "Shuko" plug for connection to
the mains. Choose the lead appropriate for your
electrical installation and discard the other.
Should neither lead be suitable, take the lead
with the U.K. plug and replace the plug with a
suitable alternative. This involves cutting off the
moulded plug, preparing the cable and
connecting to the rewireable plug in accordance
with its instructions.
IT IS IMPORTANT THAT THIS OPERATION
SHOULD ONLY BE UNDERTAKEN BY A
QUALIFIED ELECTRICIAN
NOTE: Refer to the equipment's rating plate to
ensure that the plug and fusing are suitable for
the voltage and wattage stated.
Model
Power
CB300
1200W
CB302
1250W
CB500
2250W
CR300
900W
CR302
950W

Publicidad