TEST PROGRAM SELECTION
By selecting the submenu "test", the list of test cycles that the autoclave performs can be
accessed:
1 - Vacuum Leak Test
2 - Helix Test 134°C
3 - Helix Test 121°C
4 - Bowie&Dick Test
5 - Quit
The selection indicator is positioned on "Vacuum Leak Test".
Select the desired program, moving using the (+) and (-) keys and confi rm with
VACUUM LEAK TEST
In order to perform this function, the chamber must be cold and without instruments.
BOWIE & DICK TEST
In order to perform this function, the chamber must be cold and without instruments.
The Bowie & Dick "pack" for checking the porous load must be placed on the lower tray.
HELIX TEST
In order to perform this function, the chamber must be cold and without instruments.
The control tool for examining the "Hollow A" load must be placed on the lower tray.
34
.