WARNING
To reduce the risk of burns, electrocution, allergy, fire or injury to persons:
1.
Close supervision is necessary when this product is used by, on, or near children or physically incapacitated individuals.
2.
Use this product only for its intended use as described in this guide.
3.
Never operate this product if
a.
It has a damaged power cord or plug.
b.
It is not working properly.
c.
It has been dropped or damaged.
d.
It has been dropped into water.
Return the product to an authorized DeVilbiss Healthcare service center for examination and repair.
4.
Keep the power cord away from heated surfaces.
5.
Never use while drowsy or asleep.
6.
Do not cover the unit or the AC to DC adapter while power is applied.
7.
Check the collection container for cracks before each use. Do not use if cracked.
8.
Universal precautions must be observed while providing care or handling equipment for persons with an infectious condition.
9.
Do not place the suction unit in a position that would make it difficult to disconnect the AC power cord.
10. When the device is used under extreme operating conditions and in the carrying case, the surface temperature of the device may exceed 73°C.
11. The suction tube shall be kept isolated from other parts of the system to confirm that it does not exceed 43°C prior to patient contact
12. The catheter to be used shall be carefully inserted to prevent strangulation/choking.
13. Don't let children play with the tubing as this may lead to strangulation.
14. Consult the physician in case of any allergic reaction due the use of the device.
15. Keep the suction pump out of the reach of children and pets and away from pests.
NOTE– The 7305D series is not factory equipped with an internal rechargeable battery; it may be purchased separately and installed by your DeVilbiss Healthcare provider.
7305P series is factory equipped with an internal rechargeable battery and all information regarding battery operation in this guide is applicable.
NOTE– All the materials used in the suction pump are Latex free.
DANGER
The DeVilbiss Suction Unit is a vacuum suction device designed for the collection of nonflammable fluid materials in medical applications only. Improper use
during medical applications can cause injury or death. For all medical applications:
1.
All suctioning should be done in strict accordance with appropriate procedures that have been established by a licensed medical authority.
2.
Some attachments or accessories may not fit the tubing supplied. All attachments or accessories should be checked prior to use to assure proper fit.
INTERNATIONAL TRAVEL
The 7305 series is equipped with an AC to DC adapter allowing operation on any AC voltage (100-240 VAC, 50/60 Hz). However the correct power cord must be used to
connect to adaptable wall power. AC power is removed by disconnecting the power cord from the AC outlet receptacle.
NOTE– Check power cord for adaptability before using.
INTRODUCTION
Your DeVilbiss Suction Unit is a compact medical suctioning device which has been designed for reliable, portable operation. Because of the small size, light weight and DC
operation, the DeVilbiss Suction Unit is ideal for providing suction in the home, in transport with optional DC cord, or if your model has an internal rechargeable battery, the unit
can be operated anywhere. Three container options give the choice between the standard disposable containers or an optional long-term reusable container. Following the
recommended operating and maintenance procedures outlined in this instruction guide will maximize the life of this product. This guide provides an overview of the suction unit
and operation. Your equipment provider should give more detailed instructions related to patient care.
Contraindications
(specific situation in which the device shall not be used)
The DeVilbiss Suction Unit should not be used for:
•
thoracic drainage
•
nasogastric suction
Intended Use Statement
The device is to be used to remove fluids from the airway or respiratory support system and infectious materials from wounds.The device creates a negative pressure
(vacuum) that draws fluids through disposable tubing that is connected to a collection bottle. The fluids are trapped in the collection bottle for proper disposal. It is for use on
the order of a physician only.
A-7305
EN - 3