Thermocouple Specification (Accessories):
• Operation Range: -40°C ~ 204°C (-40°F ~ 399°F).
• Accuracy: ±0.75% Reading or ±2.2°C (1.4°F), whichever is greater.
Thermistor Sensor Temperature (NTC):
• Measurement Range: 0°C to 60°C, 32°F to 140°F.
• Resolution: 0.1°C, 0.1°F.
• Accuracy: ±2°C for 0°C to 10°C.
±0.5°C for 10°C to 45°C.
±2°C for 45°C to 60°C.
±4°F for 32°F to 50°F.
±1°F for 50°F to 113°F.
±4°F for 113°F to 140°F.
Relative Humidity:
• Measurement Range: 0% to 100% RH
• Resolution: 0.1%
• Accuracy: ±2.5% at 25°C for 10% to 90% RH.
±5% at 25°C for 0% to 10% RH, 90% to 100% RH.
• Response Time: 60 seconds (for 90% of the total range).
• Sensor Hysteresis (Sensitivity for RH change from 10% to 90% to
10%): 1% RH.
• Temperature Coefficient: 0.1 time the applicable accuracy °C for
0°C to 18°C and for 28°C to 50°C (32°F to 64°F and for 82°F to
122°F).
Wireless Features:
• Frequency Range: 910~920MHz
• Current Consumption: Less than 1mA
• Transmitting Distance: 25 Meters without obstacle.
25