Changing Display From Meters To Feet; Battery; High Voltage Protection - Fluke Networks MICROSCANNER Guia Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 102
ENGLISH - 6
MICROSCANNER

Changing Display from Meters to Feet

During Calibration you will be able to switch the
displayed length from meters to feet by simply
pressing the MODE key.
Press the ON/OFF key once the desired cable
length or NVP is displayed to terminate 'Calibrate'
mode and store the new calibration factor.
MICROSCANNER will use it for future length
measurements until another calibration is
performed.

Battery

MICROSCANNER requires a 9 Volt Alkaline
battery. The Battery icon is displayed on the
screen when MICROSCANNER detects a low
battery condition.
Using MICROSCANNER with a low battery may
effect the test accuracy.
If MICROSCANNER is stored for more than one
month, the battery should be removed.
Note: MICROSCANNER will not function properly
with a 9 Volt Carbon Battery.

High Voltage Protection

MICROSCANNER is designed to withstand input
voltage conditions that arise from normal
telephony applications such as 48 VDC at less
than 80 ma or 24 VAC used to power many
keysets. Tests cannot be performed when hazard
conditions exist on the inputs.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Microscanner pro

Tabla de contenido