Connect the short silicon tube, with antibacterial filter 8, to the suction connector 5.
The other tube 4, with one end connected to the filter must be connected with the other end of jar's lid
connector 5 where has been fixed the red float (security float).
When the 90% of the volume of the jar is reached there is the activation of the security float (the float
close the aspiration connector on the jar) to avoid liquid penetration inside the device.
The device must be used on a plan of horizontal operation.
Connect the long silicon tube 1 to the other jar's lid connector 3.
Connect the other end of the long silicon tube 1 to the probe plastic connector then connect the
suction probe 6 to it.
Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and
an rapid reaching of the functionality vacuum) then rescrew the lid on the jar correctly.
Connect the power cord to the device then connect the plug to the electrical mains supply.
Push switch 7 on position I to start suction
Once finished push switch on O position and unplug.