Specifications

16.0 Specifications, Control Unit

Enclosure Dimensions
Weight
Power Cord
Overcurrent Protection
Input
Operating Ambient
Temperature Range
Classification
Electromagnetic
Compatibility
Air Express LAL Control Unit (C1500)
Detachable 14' , #18 AWG minimum, with ground wire
Class I grounded equipment not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide.
IPXO, enclosed equipment without protection against ingress of water.
CISPR 11/EN 55011, Class B, Group 1 ISM equipment
10" x 12" x 4-1/2"
(25 cm x 30 cm x 11 cm)
9.5 pounds (4.3 kg)
Primary
Two 32mm x 6.3mm fuses
5A, 250V, T, L
Secondary
One 32mm x 6.3mm fuses
0.15A, 250V,T, L
120 VAC, 60Hz, 5A
60 to 90°F (16 to 32.2°C)
Type BF equipment
UL 60601-1, CSA C22.2 NO.601.1
Continuous operation
IEC 60601-1-2:2001
9
Air Express
LAL

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

C1500M1500