Olympus OEV143 Instrucciones De Uso página 3

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Important safeguards/notices for use in the
medical environments
1. All the equipments connected to this unit shall be
certified according to Standard IEC601-1, IEC950,
IEC65 or other IEC/ISO Standards applicable to the
equipments.
2. When this unit is used together with other equipment in
the patient area*, the equipment shall be either powered
by an isolation transformer or connected via an
additional protective earth terminal to system ground
unless it is certified according to Standard IEC601-1
and IEC601-1-1.
* Patient Area
3. The leakage current could increase when connected to
other equipment.
4. The operator should take precautions to avoid touching
the rear panel input and output circuitry and the patient
at the same time.
5. Model OEV143/OEV203 is a video monitor intended
for use in a medical environment to display video
pictures from EVIS video system center or other video
system.
Warning on power connection
Use a proper power cord for your local power supply.
Plug type
Female end
Cord type
Minimum cord set rating
Safety approval
*Note: Grounding reliability can only be achieved when the equipment is connected to
an equivalent receptacle marked 'Hospital Only' or 'Hospital Grade'.
United State
Canada
HOSPITAL GRADE*
Min.Type SJT
Min.18 AWG
10A/125V
10A/125V
UL Listed
CSA
Symbols on the unit
Symbol
Location
Front panel
Rear panel
Rear panel
Rear panel
Rear panel
Continental Europe
LP-34A or LP-34
LS-60
H05VV-F
10A/250V
VDE
This symbol indicates
Main power switch.
Press to turn the monitor
on or off.
The equipotential terminal
which brings the various
parts of a system to the
same potential.
Functional earth terminal
Alternating current
Attention, consult
ACCOMPANYING
DOCUMENTS
Japan
VM1050
VM1010
PVCTF
12A/125V
DENAN-HO
3
EN

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Oev203

Tabla de contenido