1. Introduction
BOD measurement with BOD EVO Sensor is based on the respirometric method that provides a direct measurement of the
oxygen consumed by microorganisms from an air or oxygen-enriched environment in a closed atmosphere, at a constant
temperature (20 °C ± 0.5) and with a soft sample stirring.

2. Assembly and installation

Upon receipt and after having removed the packaging, please check the integrity of the instrument. The box includes:
● BOD EVO Sensor
BOD EVO Sensor is delivered with 2 lithium batteries installed: remove the insulating strip from the rear without opening it.

3. Operating controls

The display of the electronic measuring unit is normally turned off reducing he consumption of internal batteries. The touch
of a key activates the display which automatically is turned off after three seconds. The key activates different functions if
the unit is performing a measurement or if it is in memory code and if the unit is (or not) in wireless mode. When the display
is turned off and no analysis is performed, BOD EVO Sensor shows the symbol
3.1 Wireless or non-wireless mode
BOD EVO Sensor is already set on the wireless mode. To switch to a non-wireless mode, keep pressed SET when the
display is off until "S.E.T." appears. When the three dots disappear:
Key SET
Pass to wireless mode ON
3.2 Description of keys (wireless mode)
Out of a measuring cycle (before starting)
Key SET with display off
Key SET with display on
Key START with display off
Key START with display on
During a measuring cycle (after starting)
Key SET
Key START
Key START kept pressed
Memory mode (it is active during a measuring cycle and also after if Reset is not performed)
Key SET
Key START
Reset
Keys SET+START kept pressed
Before starting a new measuring cycle, it is requested to reset the stored values.
3.3 Description of keys (non-wireless mode)
Out of a measuring cycle (before starting)
Key SET with display off
Key SET with display on
Key START with display off
Key START with display on
During a measuring cycle (after starting)
Key SET
Key START
Key START kept pressed
Memory mode (it is active during a measuring cycle and also after if Reset is not performed)
Key SET
Key START
If no values are stored, the display shows three points after a prolonged pressing of key START.
Reset
Keys SET+START kept pressed
Before starting a new measuring cycle, it is requested to reset the stored values.
NOTE: before beginning an additional BOD EVO Sensor, it's necessary to wait that the display shows "An".
3.4 Last data recovery
If during an analysis the Wireless DataBox™ looses some data, it's possible to retrieve the data for the last seven samplings
by pressing SET twice: in this way data will be automatically sent to the Wireless DataBox™.
3.5 Scale selection
Scale
A: 0 ÷ 999 mg O
B: 0 ÷ 600 mg O
C: 0 ÷ 250 mg O
D: 0 ÷ 90 mg O
● Alkali holder
Wireless symbol
Wireless symbol
Wireless symbol
The cycle of measurement is started
The scale in use is shown
The current measure is shown
The memory mode is entered
The stored value is shown
The order number of stored value is increased
Stops the measurement cycle, deletes the stored values and shows the scale
The scale is shown
The scale is changed
The scale is shown
The cycle of measurement is started
The scale is shown
The current measure is shown
The memory mode is entered
The stored value is shown
The order number of stored value is increased
Measurement cycle stopped, stored values deleted and the scale is shown
Volume of sample
/l
2
/l
2
/l
2
/l
2
Key START
is shown
is shown
is shown
(also after the 5-day period if Reset is not performed)
(also after the 5-day period if Reset is not performed)
(also after the 5-day period if Reset is not performed)
(also after the 5-day period if Reset is not performed)
100 ml
150 ml
250 ml
400 ml
3
● Instruction manual
by pressing any keys, in wireless mode.
Pass to wireless mode OFF
The scale is related to the
expected BOD value.
The possibility of measuring
diluted samples extends the range.
EN